Beck, J A; Budgell, G J; Roberts, D A; Evans, P M.
Afiliação
  • Beck JA; North Western Medical Physics, Christie Hospital NHS Foundation Trust, Withington, Manchester M20 4BX, United Kingdom. james.beck@physics.cr.man.ac.uk
Med Phys ; 36(5): 1859-66, 2009 May.
Article em En | MEDLINE | ID: mdl-19544805
ABSTRACT
An amorphous silicon EPID has been investigated to determine whether it is capable of quality control constancy measurements for linear accelerator electron beams. The EPID grayscale response was found to be extremely linear with dose over a wide dose range and, more specifically, for exposures of 95-100 MU. Small discrepancies of up to 0.8% in linearity were found at 6 MeV (8-15 MeV showed better agreement). The shape of the beam profile was found to be significantly altered by scatter in air over the approximately 60 cm gap between the end of the applicator and the EPID. Nevertheless, relative changes in EPID-measured profile flatness and symmetry were linearly related to changes in these parameters at 95 cm focus to surface distance (FSD) measured using a 2D diode array. Similar results were obtained at 90 degrees and 270 degrees gantry angles. Six months of daily images were acquired and analyzed to determine whether the device is suitable as a constancy checker. EPID output measurements agreed well with daily ion chamber measurements, with a 0.8% standard deviation in the difference between the two measurement sets. When compared to weekly parallel plate chamber measurements, this figure dropped to 0.5%. A Monte Carlo (MC) model of the EPID was created and demonstrated excellent agreement between MC-calculated profiles in water and the EPID at 95 and 157 cm FSD. Good agreement was also found with measured EPID profiles, demonstrating that the EPID provides an accurate measurement of electron profiles. The EPID was thus shown to be an effective method for performing electron beam daily constancy checks.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Silício / Ecrans Intensificadores para Raios X / Radioterapia Conformacional Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Reino Unido País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Garantia da Qualidade dos Cuidados de Saúde / Radiometria / Silício / Ecrans Intensificadores para Raios X / Radioterapia Conformacional Tipo de estudo: Diagnostic_studies / Prognostic_studies Idioma: En Revista: Med Phys Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Reino Unido País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA