Design and development of a portable gamma radiation monitor.
Munir, M; Ahmad, N; Sohail, S; Naveed, R A; Rafiq, M Q; Khalid, M.
Afiliação
  • Munir M; Health Physics Division, Directorate of Systems and Services, Pakistan Institute of Nuclear Science and Technology (PINSTECH), Nilore, Islamabad 45650, Pakistan. mmunirali@yahoo.com
Rev Sci Instrum ; 80(7): 073101, 2009 Jul.
Article em En | MEDLINE | ID: mdl-19655937
ABSTRACT
A portable gamma radiation monitor has been designed and developed. The monitor can be used effectively in the dose range from 0.07 to 500 mGy/h due to gamma rays of energy greater than 65 keV. The monitor overestimated radiation doses and the uncertainty in the measured dose rate has been found to be < or = 30%. The response of the monitor can be considered isotropic within an acceptable error of +/-30%. Provision has also been added to use the monitor as an installed radiation monitor. In installed mode, it can be operated from a remote location up to 1 km and the timing history can be stored on a personal computer.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitoramento de Radiação / Raios gama Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Paquistão País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Monitoramento de Radiação / Raios gama Idioma: En Revista: Rev Sci Instrum Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Paquistão País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA