Analytical performance evaluation of the Elecsys® Troponin T Gen 5 STAT assay.
Clin Chim Acta ; 495: 522-528, 2019 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-31152695
ABSTRACT

BACKGROUND:

We report the analytical performance of the Elecsys® Troponin T Gen 5 STAT (TnT Gen 5 STAT; Roche Diagnostics) assay.

METHODS:

Measuring limits/ranges were determined in lithium-heparin plasma samples per Clinical and Laboratory Standards Institute (CLSI) EP17-A2. Precision was evaluated per CLSI EP05-A2 using lithium-heparin plasma/quality control samples on cobas e 411/cobas e 601 analyzers; two duplicated runs per day for 21 days (n = 84). Cross-reactivity with other troponin forms and interference from endogenous substances/drugs was tested; recovery criterion for no cross-reactivity was within ±10%.

RESULTS:

Coefficients of variation (CV) for repeatability/intermediate precision were 0.7-5.6%/1.4-10.3% (cobas e 411; mean cardiac troponin T [cTnT]: 7.3-9341 ng/L) and 0.7-3.0%/1.5-6.4% (cobas e 601; mean cTnT: 7.4-9455 ng/L). There was no cross-reactivity with skeletal muscle troponin T (≤ 10,000 ng/L), skeletal muscle troponin I (≤ 100,000 ng/L), cardiac troponin I (≤ 10,000 ng/L), or human troponin C (≤ 80,000 ng/L). No interference was observed with biotin (≤ 20 ng/mL) or 34 drugs.

CONCLUSION:

The TnT Gen 5 STAT assay demonstrated a CV of <10% at the 99th percentile upper reference limit, meeting precision requirements (Fourth Universal Definition of Myocardial Infarction) for high-sensitivity troponin assays.
Assuntos

Similares

MEDLINE

...
LILACS

LIS

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: MEDLINE Assunto principal: Análise Química do Sangue / Imunoensaio / Troponina T Aspecto clínico: Diagnóstico Limite: Humanos Idioma: Inglês Revista: Clin Chim Acta Ano de publicação: 2019 Tipo de documento: Artigo