Ex-PRESS implantation for different types of glaucoma.
Altinel, Meltem Guzin; Kanra, Ayse Yagmur; Karadag, Remzi; Bayramlar, Huseyin.
Afiliação
  • Altinel MG; Department of Ophthalmology, Fatih Sultan Mehmet Training and Research Hospital, Icerenkoy Atasehir 34752, Istanbul, Turkey.
  • Kanra AY; Dunyagoz Hospitals Group, Sultanbeyli 34337, Istanbul, Turkey.
  • Karadag R; Department of Ophthalmology, Istanbul Medeniyet University School of Medicine, Goztepe Research and Training Hospital, Goztepe 34000, Istanbul, Turkey.
  • Bayramlar H; Dunyagoz Hospitals Group, Sultanbeyli 34337, Istanbul, Turkey.
Int J Ophthalmol ; 12(8): 1290-1297, 2019.
Article em En | MEDLINE | ID: mdl-31456919
ABSTRACT

AIM:

To evaluate the clinical results, efficacy and safety of Ex-PRESS P200 glaucoma shunt implant in different types of medically uncontrolled glaucoma.

METHODS:

The study included 31 eyes of 31 patients that were unresponsive to medical antiglaucomatous therapy in whom Ex-PRESS P200 glaucoma shunt implantation was performed. The demographic characteristics of the patients, type of glaucoma, complete ocular examination results, number of antiglaucomatous drugs before and after surgery, early and late complications of surgery, additional surgical and nonsurgical medical interventions, and success rates were investigated from the patients' files, retrospectively.

RESULTS:

The mean postoperative follow-up time was 16.4±7.5mo. The preoperative mean corrected intraocular pressure (IOP) was 28.7±10.3 mm Hg and postoperative mean corrected IOP was 15.3±5.2 mm Hg (P<0.05) at the last visit. The mean IOP reduction was 39.9% when the preoperative and postoperative values of the last visits were compared. The average number of antiglaucomatous drug use decreased from 3.9±0.3 to 1.7±1.7 postoperatively (P<0.05). The use of antiglaucomatous medications at the last visit was more than in other studies in the literature. The most common complication was conjunctival leakage, which was seen in 7 patients. Other early complications were iris touch, intravitreal hemorrhage, hyphema, choroidal effusion, early transient hypotonia and corneal edema. One of the late complications was endophthalmitis which was seen in one case 6mo after the operation, and the other late complication was opacification of the cornea in one patient. Twelve additional surgical operations associated to Ex-PRESS surgery and 3 bleb needling have done. At the last visit, the complete success rate was 32.3% and the qualified success rate was 77.5% in all patients.

CONCLUSION:

Ex-PRESS P200 glaucoma shunt implantation may be an effective procedure for medically uncontrolled glaucoma with significantly lower use of antiglaucomatous medications.