Your browser doesn't support javascript.
loading
Validation of a Parvovirus B19 NAT Assay for Screening of Umbilical Cord Blood for Allogenic Hematopoietic Stem Cell Donation.
Steininger, Philipp; Korn, Klaus; Hackstein, Holger; Strasser, Erwin F.
Afiliação
  • Steininger P; Institute of Clinical and Molecular Virology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Korn K; Institute of Clinical and Molecular Virology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Hackstein H; Department of Transfusion Medicine and Haemostaseology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
  • Strasser EF; Division of Transfusion Medicine, Cell Therapeutics and Haemostaseology, University Hospital, LMU Munich, Munich, Germany.
Transfus Med Hemother ; 51(1): 48-51, 2024 Feb.
Article em En | MEDLINE | ID: mdl-38314242
ABSTRACT

Introduction:

Parvovirus B19 transmitted by umbilical cord blood (UCB) products may cause severe disease in allogenic hematopoietic stem cell transplant recipients. Thus, commercially available nucleic acid test (NAT) assays for highly sensitive detection of parvovirus B19 DNA validated for the specimen cord blood plasma (CBP) are required to avoid parvovirus B19 transmission by umbilical hematopoietic stem cell preparations.

Methods:

The multiplex cobas DPX NAT assay was validated for detection of parvovirus B19 DNA in CBP derived from citrate anticoagulated UCB units which have been processed by the Rubinstein method. In total, 363 retained CBP samples pretested negative for parvovirus B19 DNA were prepared for analyzing sensitivity, specificity, and interference of that NAT assay. The 3rd WHO International Standard for parvovirus B19 DNA was used for determining the 95% limit of detection (LOD95) by probit analysis.

Results:

The validation of the parvovirus B19 NAT assay for CBP demonstrated high sensitivity, specificity, intra- and inter-assay precision. Dilution series and replicate analyses showed a high linearity of the assay with a coefficient of determination above 0.99 and revealed a LOD95 of 17 International Units (IU)/mL (95% confidence interval, 14-44 IU/mL) for parvovirus B19 DNA in CBP samples.

Conclusion:

The validation of a commercially available parvovirus B19 NAT assay for the specimen CBP demonstrated a high assay performance fulfilling German guidelines and international regulations.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Screening_studies Idioma: En Revista: Transfus Med Hemother Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Suíça

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Screening_studies Idioma: En Revista: Transfus Med Hemother Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha País de publicação: Suíça