Sub-ppt level detection of carbon monoxide and nitrous oxide enabled by mid-infrared doubly resonant photoacoustic spectroscopy.
Opt Lett ; 49(13): 3648-3651, 2024 Jul 01.
Article em En | MEDLINE | ID: mdl-38950231
ABSTRACT
We report highly sensitive detection of carbon monoxide (CO) and nitrous oxide (N2O) using doubly resonant photoacoustic spectroscopy paired with a quantum cascade laser (QCL) at 4.57 µm. The butterfly-packaged QCL is used to exploit the CO absorption line at 2190.02 cm-1 and the N2O absorption line at 2191.42 cm-1 by scanning the injection current. Leveraging the simultaneous acoustic and optical resonances and adopting a lower photoacoustic detection frequency, we achieve a minimum detection limit of 0.85 part-per-trillion (ppt) for CO over the 500 s averaging time, and 0.7 ppt for N2O over the 200 s averaging time. Our approach demonstrates record sensitivity for CO and N2O detection compared to state-of-the-art optical gas sensors.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Opt Lett Ano de publicação: 2024 Tipo de documento: Article País de publicação: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA