Study of Key Electrochemical Problems in Transdermal Noninvasive Glucose Monitoring / 分析化学
Chinese Journal of Analytical Chemistry
; (12): 1566-1571, 2009.
Article
em Zh
| WPRIM
| ID: wpr-404440
Biblioteca responsável:
WPRO
ABSTRACT
Several key problems were analysed for the transdermal noninvasive glucose monitoring. A modified calibration equation was proposed for the high-sensitivity glucose biosensor due to its narrow linear range. The new equation has increased the sensors' linear range by 20 times. A new diffuse model was constructed for the electrode system of glucose sensor,aiming at the unique "finite space" electrochemical problem in trans-dermal technique. In addition,electrode masks were utilized to solve the problem of electrode loss in longtime glucose monitoring. In animal in-vivo experiments,70.1 % of the noninvasive glucose data points were clinically accurate,while the remains were clinically acceptable. All solutions mentioned above were based on both theoretical analysis and experimental validation,promoting the realization and optimization of transdermal noninvasive glucose monitoring techniques.
Texto completo:
1
Base de dados:
WPRIM
Idioma:
Zh
Revista:
Chinese Journal of Analytical Chemistry
Ano de publicação:
2009
Tipo de documento:
Article