1.
Ann Biomed Eng ; 52(5): 1403-1414, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38402315

RESUMO

During follow-up of patients treated with WEB devices, shape changes have been observed. The quantitative three-dimensional measurement of the WEB shape modification (WSM) would offer useful information to be studied in association with the anatomical results and try to better understand mechanisms implicated in this modification phenomenon. We present a methodology to quantify the morphology and position of the WEB device in relation to the vascular anatomy. Three-dimensional rotational angiography (3DRA) images of seven aneurysms patients treated with WEBs were used, which also accompanied by a post-treatment 3DRA image and a follow-up 3DRA image. The device was manually segmented, obtaining the 3D models after treatment and at the follow-up. Volume, surface area, height, maximum diameter and WSM ratio of both surfaces were calculated. Position changes were evaluated measuring WEB axis and relative position between post-treatment and follow-up. Changes in WEB volume and surface area were observed with a mean modification of - 5.04 % ( ± 14.19 ) and - 1.68 % ( ± 8.29 ) , respectively. The positional variables also showed differences, mean change of device axis direction was 26.25 % ( ± 24.09 ) and mean change of distance l b was 5.87 % ( ± 10.59 ) . Inter-observer and intra-observer variability analyses did not show differences (ANOVA p > 0.05 ). This methodology allows quantifying the morphological and position changes suffered by the WEB device after treatment, offering new information to be studied in relation to the occurrence of WEB shape modification.


Assuntos
Embolização Terapêutica , Procedimentos Endovasculares , Aneurisma Intracraniano , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/terapia , Variações Dependentes do Observador , Resultado do Tratamento , Estudos Retrospectivos , Angiografia Cerebral/métodos
2.
Artigo em Inglês | MEDLINE | ID: mdl-39160330

RESUMO

PURPOSE: To introduce a computational tool for peri-interventional intracranial aneurysm treatment guidance that maps preoperative planning information from simulation onto real-time X-Ray imaging. METHODS: Preoperatively, multiple flow diverter (FD) devices are simulated based on the 3D mesh of the vessel to treat, to choose the optimal size and location. In the peri-operative stage, this 3D information is aligned and mapped to the continuous 2D-X-Ray scan feed from the operating room. The current flow diverter position in the 3D model is estimated by automatically detecting the distal FD marker locations and mapping them to the treated vessel. This allows to visually assess the possible outcome of releasing the device at the current position, and compare it with the one chosen pre-operatively. RESULTS: The full pipeline was validated using retrospectively collected biplane images from four different patients (5 3D-DSA datasets in total). The distal FD marker detector obtained an average F1-score of 0.67 ( ± 0.224 ) in 412 2D-X-Ray scans. After aligning 3D-DSA + 2D-X-Ray datasets, the average difference between simulated and deployed positions was 0.832 mm ( ± 0.521 mm). Finally, we qualitatively show that the proposed approach is able to display the current location of the FD compared to their pre-operatively planned position. CONCLUSIONS: The proposed method allows to support the FD deployment procedure by merging and presenting preoperative simulation information to the interventionists, aiding them to make more accurate and less risky decisions.

3.
Int J Comput Assist Radiol Surg ; 16(10): 1663-1673, 2021 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-34195929

RESUMO

MOTIVATION: Intra-saccular devices (ID), developed for the treatment of bifurcation aneurysms, offer new alternatives for treating complex terminal and bifurcation aneurysms. In this work, a complete workflow going from medical images to post-treatment CFD analysis is described and used in the assessment of a concrete clinical problem. MATERIALS AND METHODS: Two different intra-saccular device sizes were virtually implanted in 3D models of the patient vasculature using the ID-Fit method. After deployment, the local porosity at the closed end of the device in contact with the blood flow was computed. This porosity was then used to produce a CFD porous medium model of the device. Velocities and wall shear stress were assessed for each model. RESULTS: Six patients treated with intra-saccular devices were included in this work. For each case, 2 different device sizes were virtually implanted and 3 CFD simulations were performed: after deployment simulation with each size and before deployment simulation (untreated). A visible reduction in velocities was observed after device implantation. Velocity and WSS reduction was statistically significant (K-S statistics, [Formula: see text]). CONCLUSIONS: Placement of different device size can lead to a partial filling of the aneurysm, either at the dome or at the neck, depending on the particular positioning by the interventionist. The methodology used in this work can have a strong clinical impact, since it provides additional information in the process of device selection using preoperative data.


Assuntos
Aneurisma Intracraniano , Simulação por Computador , Hemodinâmica , Humanos , Aneurisma Intracraniano/diagnóstico por imagem , Aneurisma Intracraniano/cirurgia , Modelos Cardiovasculares , Estresse Mecânico
4.
J Biomech ; 110: 109945, 2020 09 18.
Artigo em Inglês | MEDLINE | ID: mdl-32827768

RESUMO

Modelling intracranial aneurysm blood flow after flow diverter treatment has proven to be of great scientific and clinical interest. One of the reasons for not having CFD as an everyday clinical tool yet is the time required to set-up such simulations plus the required computational time. The speed-up of these simulations can have a considerable impact during treatment planning and device selection. Modelling flow diverters as a porous medium (PM) can considerably improve the computational time. Many models have been presented in literature, but quantitative comparisons between models are scarce. In this study, the untreated case, the explicit definition of the flow diverter wires as no-slip boundary condition and five different porous medium models were chosen for comparison, and evaluated on intracranial aneurysm of 14 patients with different shapes, sizes, and locations. CFD simulations were made using finite volume method on steady flow conditions. Velocities, kinetic energy, wall shear stress, and computational time were assessed for each model. Then, all models are compared against the no-slip boundary condition using non parametric Kolmogorov-Smirnov test. The model with least performance showed a mean K-S statistic of 0.31 and deviance of 0.2, while the model with best values always gave K-S statistics below 0.2. Kinetic energy between PM models varied between an over estimation of 218.3% and an under estimation of 73.06%. Also, speedups were between 4.75x and 5.3x (stdev: 0.38x and 0.15x) when using PM models. Flow diverters can be simulated with PM with a good agreement to standard CFD simulations were FD wires are represented with no-slip boundary condition in less than a quarter of the time. Best results were obtained on PM models based on geometrical properties, in particular, when using a heterogeneous medium based on equations for flat rhomboidal wire frames.


Assuntos
Aneurisma Intracraniano , Stents , Simulação por Computador , Hemodinâmica , Humanos , Modelos Cardiovasculares , Porosidade , Estresse Mecânico
5.
Int J Numer Method Biomed Eng ; 34(12): e3145, 2018 12.
Artigo em Inglês | MEDLINE | ID: mdl-30152120

RESUMO

In computational fluid dynamics, there is a high interest in modeling flow diverter stents as porous media due to its reduced computational loads. One of the main difficulties of such models is proper parameter setup. Most authors assume flow diverter's wire screen as an isotropic and homogeneous medium, while others proposes anisotropic configurations, yet very little is discussed about the effect of these assumptions on model's accuracy. In this paper, we compare the effect of different models on hemodynamics in relation to their parameters. The fidelity and efficiency of the different models to capture wire screen effect on fluid flow are quantitatively analyzed and compared.


Assuntos
Simulação por Computador , Hemodinâmica , Modelos Cardiovasculares , Desenho de Prótese , Stents , Humanos , Porosidade
SELEÇÃO DE REFERÊNCIAS