This paper presents the gamma-ray detection performance of the newly developed MAPD-3NM-II type SiPM sensor array (4 [Formula: see text] 4) with [Formula: see text](Ce) scintillator. The gamma-ray spectra of various sources have been measured in the energy range from 26 keV up to 1332 keV. The newly developed array based on MAPD-3NM-II sensors proved [Formula: see text] 22% enhancement in energy resolution in comparison to the former MAPD-3NM-I based array. The energy resolution of 662 keV gamma-rays measured by MAPD-3NM-II was 3.3% while clearly surpassing 4.25% resolution of MAPD-3NM-I predecessor. The enhancement is related to the high PDE of the new MAPD-3NM-II. Obtained results show that the new MAPD-3NM-II demonstrated good energy resolution and linearity in the studied energy region. The energy resolution of the new detector developed based on MAPD-3NM-II was better than all previous products of MAPD.