J Cardiovasc Electrophysiol ; 30(3): 320-325, 2019 03.
Artigo em Inglês | MEDLINE | ID: mdl-30556201

RESUMO

BACKGROUND: Implantable cardioverter defibrillator (ICD) lead failures occur at higher rates in pediatric and congenital heart disease (CHD) patients. OBJECTIVE: To determine the rate and timing of Riata lead failure in pediatric and CHD patients. METHODS: This was a retrospective, multicenter cohort study of pediatric patients and adults with CHD with implantation of a Riata or Riata ST lead between 2002 and 2009. The prevalence and timing of electrical failure and conductor coil externalization (CCE) were determined. RESULTS: Fifty-eight patients and 63 leads from seven centers were included. Median (interquartile range [IQR]) age at implant was 14.4 (11.5-18.7) years and median follow-up was 8.7 (7.3-11.1) years. The underlying diagnosis was a primary arrhythmia disorder in 45%, cardiomyopathy in 31%, and CHD in 28% of patients. Electrical failure occurred in 43% and CCE in 16% of leads at median lead ages of 4.7 (3.4-7.5) and 4.3 (3.9-7.0) years, respectively. Median lead survival free from electrical failure or CCE was 7.9 (95% confidence interval, 5.8-10.0) years. Forty-one percent of leads were functional at the end of the follow-up period, and 33% were extracted with a complication rate of 5%. CONCLUSIONS: The rate of Riata lead electrical failure was high in children and patients with CHD, while the rate of CCE was comparable with published data. Counseling on lead management should factor in the high rate of electrical failure with considerations for elective replacement.


Assuntos
Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Cardiopatias Congênitas/terapia , Falha de Prótese , Adolescente , Fatores Etários , Canadá , Criança , Pré-Escolar , República Tcheca , Remoção de Dispositivo , Cardioversão Elétrica/efeitos adversos , Feminino , Cardiopatias Congênitas/diagnóstico , Cardiopatias Congênitas/fisiopatologia , Humanos , Masculino , Intervalo Livre de Progressão , Desenho de Prótese , Estudos Retrospectivos , Medição de Risco , Fatores de Risco , Fatores de Tempo , Estados Unidos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA