Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 7 de 7
Filtrar
Mais filtros

Base de dados
Tipo de documento
Intervalo de ano de publicação
1.
J Mech Behav Biomed Mater ; 144: 105951, 2023 08.
Artigo em Inglês | MEDLINE | ID: mdl-37295386

RESUMO

In the present work, we test four thin coatings for titanium implants, namely, bioglass, GB14, Beta-Tricalciumphosphate (ß-TCP) and hydroxyapatite (HA) with and without incorporated copper ions for their osteointegrative capacity. A rabbit drill hole model for time intervals up to 24 weeks was used in this study. Implant fixation was evaluated by measuring shear strength of the implant/bone interface. Quantitative histological analysis was performed for the measurements of bone contact area. Implants with and without copper ions were compared after 24 weeks. Thin coatings of GB14, HA or TCP on titanium implants demonstrated high shear strength during the entire test period of up to 24 weeks. Results confirmed osteointegrative properties of the coatings and did not reveal any negative effect of copper ions on osteointegration. The integration of copper in degradable osteoconductive coatings with a thickness of approx. 20 µm represents a promising method of achieving antibacterial shielding during the entire period of bone healing while at the same time improving osteointegration of the implants.


Assuntos
Cobre , Durapatita , Animais , Coelhos , Titânio , Cerâmica , Propriedades de Superfície , Materiais Revestidos Biocompatíveis/farmacologia , Osseointegração
2.
Materials (Basel) ; 16(18)2023 Sep 19.
Artigo em Inglês | MEDLINE | ID: mdl-37763556

RESUMO

High-velocity oxygen fuel (HVOF) spraying is a promising technique for depositing protective coatings. The performances of HVOF-sprayed coatings are affected by in-flight particle properties, such as temperature and velocity, that are controlled by the spraying parameters. However, obtaining the desired coatings through experimental methods alone is challenging, owing to the complex physical and chemical processes involved in the HVOF approach. Compared with traditional experimental methods, a novel method for optimizing and predicting coating performance is presented herein; this method involves combining machine learning techniques with thermal spray technology. Herein, we firstly introduce physics-informed neural networks (PINNs) and convolutional neural networks (CNNs) to address the overfitting problem in small-sample algorithms and then apply the algorithms to HVOF processes and HVOF-sprayed coatings. We proposed the PINN and CNN hierarchical neural network to establish prediction models for the in-flight particle properties and performances of NiCr-Cr3C2 coatings (e.g., porosity, microhardness, and wear rate). Additionally, a random forest model is used to evaluate the relative importance of the effect of the spraying parameters on the properties of in-flight particles and coating performance. We find that the particle temperature and velocity as well as the coating performances (porosity, wear resistance, and microhardness) can be predicted with up to 99% accuracy and that the spraying distance and velocity of in-flight particles exert the most substantial effects on the in-flight particle properties and coating performance, respectively. This study can serve as a theoretical reference for the development of intelligent HVOF systems in the future.

3.
Materials (Basel) ; 14(1)2020 Dec 31.
Artigo em Inglês | MEDLINE | ID: mdl-33396431

RESUMO

Bone defects introduced by accidents or diseases are very painful for the patient and their treatment leads to high expenses for the healthcare systems. When a bone defect reaches a critical size, the body is not able to restore this defect by itself. In this case a bone graft is required, either an autologous one taken from the patient or an artificial one made of a bioceramic material such as calcium phosphate. In this study ß-tricalcium phosphate (ß-TCP) was dispersed in a polymer matrix containing poly(lactic acid) (PLA) and poly(ethylene glycole) (PEG). These compounds were extruded to filaments, which were used for 3D printing of cylindrical scaffolds via Fused Deposition of Ceramics (FDC) technique. After shaping, the printed parts were debindered and sintered. The components combined macro- and micropores with a pore size of 1 mm and 0.01 mm, respectively, which are considered beneficial for bone healing. The compressive strength of sintered cylindrical scaffolds exceeded 72 MPa at an open porosity of 35%. The FDC approach seems promising for manufacturing patient specific bioceramic bone grafts.

4.
Materials (Basel) ; 13(3)2020 Jan 30.
Artigo em Inglês | MEDLINE | ID: mdl-32019258

RESUMO

Thermal spraying of metal materials is one of the key applications of this technology in industry for over a hundred years. The variety of metal-based feedstocks (powders and wires) used for thermal spray is incredibly large and utilization covers abrasion and corrosion protection, as well as tribological and electrical applications. Spraying metals using suspension- or precursor-based thermal spray methods is a relatively new and unusual approach. This publication deals with three metal types, a NiCr 80/20, copper (Cu), and silver (Ag), sprayed as fine-grained powders dispersed in aqueous solvent. Suspensions were sprayed by means of high-velocity suspension spraying (HVSFS) employing a modified TopGun system. The aim was to prepare thin and dense metal coatings (10-70 µm) and to evaluate the process limits regarding the oxygen content of the coatings. In case of Cu and Ag, possible applications demand high purity with low oxidation of the coating to achieve for instance a high electrical conductivity or catalytic activity. For NiCr however, it was found that coatings with a fine dispersion of oxides can be usable for applications where a tunable resistivity is in demand. The paper describes the suspension preparation and presents results of spray experiments performed on metal substrates. Results are evaluated with respect to the phase composition and the achieved coating morphology. It turns out that the oxidation content and spray efficiency is strongly controlled by the oxygen fuel ratio and spray distance.

5.
Materials (Basel) ; 12(21)2019 Oct 25.
Artigo em Inglês | MEDLINE | ID: mdl-31731410

RESUMO

One of the most common causes of implant failure is aseptic prosthesis loosening. Another frequent complication after prosthesis implant is the microbial colonization of the prosthesis surface, which often leads to a replacement of the prosthesis. One approach to reduce these complications is the application of bioactive substances to implant surfaces. Both an antibiotic prophylaxis and a faster osteointegration can be obtained by incorporation of bactericidal active metals in degradable calcium phosphate (CaP) coatings. In this study, thin degradable calcium phosphate ceramic coatings doped with silver (Ag), copper (Cu), and bismuth (Bi) on a titanium substrate were prepared with the aid of the high-velocity suspension flame spraying (HVSFS) coating process. To characterize the samples surface roughness, brightfield microscopy of the coatings, X-ray diffraction (XRD)-analysis for definition of the phase composition of the layers, Raman spectroscopy for determination of the phase composition of the contained metals, element-mapping for Cu-content verification, release kinetics for detection of metal ions and ceramic components of the coatings were carried out. The aim of this study was to evaluate in vitro biocompatibility and antimicrobial activity of the coatings. For biocompatibility testing, growth experiments were performed using the cell culture line MG-63. Cell viability was investigated by Giemsa staining and live/dead assay. The WST-1 kit was used to quantify cell proliferation and vitality in vitro and the lactate dehydrogenase (LDH) kit to quantify cytotoxicity. The formation of hydroxyapatite crystals in simulated body fluid was investigated to predict bioactivity in vivo. The Safe Airborne Antibacterial Assay with Staphylococcus aureus (S. aureus) was used for antimicrobial testing. The results showed good biocompatibility of all the metal doped CaP coatings, furthermore Cu and Ag doped layers showed significant antibacterial effects against S. aureus.

6.
Dent Mater ; 31(8): e157-67, 2015 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-25981907

RESUMO

OBJECTIVES: Wettability is increasingly considered to be an important factor determining biological responses to implant materials. In this context, the purpose of this study was to compare the dynamic wettability of dental implants made from different bulk materials and modified by different surface modifications, and to analyze the respective changes of wettability upon irradiating these implants by UV-A or UV-C light. METHODS: Four original screw-type implants were investigated: One grit-blasted/acid-etched and one anodically oxidized titanium, one zirconia and one polyetheretherketone implant. Additionally, experimental, screwless, machined titanium cylinders were included in the study. Part of that cylinders and of blasted/etched implants were further modified by a magnetron-sputtered photocatalytic anatase thin film. Scanning electron microscopy was used to investigate the surface micro- and nanostructures. Samples were treated by UV-A (382nm, 25mWcm(-2)) and UV-C (260nm, 15mWcm(-2)) for entire 40min, respectively, and their wettability was quantified by dynamic contact angle (CA) analysis from multi-loop Wilhelmy experiments. RESULTS: All implants are characterized by submicron- and nanosized surface features. Unexposed implants were hydrophobic (CA>90°). Upon UV-A, solely the implants with anatase coating became superhydrophilic (CA<5°). Upon UV-C, the blasted/etched implants turned superhydrophilic, the anodized titanium and the zirconia implants were considerably (CA=34° and 27°, respectively) and the PEEK implants slightly (CA=79°) hydrophilized. SIGNIFICANCE: The wettability of implant surfaces can be improved by UV irradiation. The efficiency of UV-A and UV-C irradiation to lower the CA by photocatalysis or photolysis, however, is strongly dependent on the specific material and surface. Thus, attempts to photofunctionalize these surfaces by irradiation is expected to result in a different pattern of bioresponses.


Assuntos
Implantes Dentários , Interações Hidrofóbicas e Hidrofílicas , Cetonas/química , Polietilenoglicóis/química , Titânio/química , Raios Ultravioleta , Zircônio/química , Benzofenonas , Teste de Materiais , Microscopia Eletrônica de Varredura , Polímeros , Propriedades de Superfície , Molhabilidade
7.
Mater Sci Eng C Mater Biol Appl ; 34: 287-303, 2014 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-24268261

RESUMO

In cementless fixation of metallic prostheses, bony ingrowth onto the implant surface is often promoted by osteoconductive plasma-sprayed hydroxyapatite coatings. The present work explores the use of the innovative High Velocity Suspension Flame Spraying (HVSFS) process to coat Ti substrates with thin homogeneous hydroxyapatite coatings. The HVSFS hydroxyapatite coatings studied were dense, 27-37µm thick, with some transverse microcracks. Lamellae were sintered together and nearly unidentifiable, unlike conventional plasma-sprayed hydroxyapatite. Crystallinities of 10%-70% were obtained, depending on the deposition parameters and the use of a TiO2 bond coat. The average hardness of layers with low (<24%) and high (70%) crystallinity was ≈3.5GPa and ≈4.5GPa respectively. The distributions of hardness values, all characterised by Weibull modulus in the 5-7 range, were narrower than that of conventional plasma-sprayed hydroxyapatite, with a Weibull modulus of ≈3.3. During soaking in simulated body fluid, glassy coatings were progressively resorbed and replaced by a new, precipitated hydroxyapatite layer, whereas coatings with 70% crystallinity were stable up to 14days of immersion. The interpretation of the precipitation behaviour was also assisted by surface charge assessments, performed through Z-potential measurements. During in vitro tests, HA coatings showed no cytotoxicity towards the SAOS-2 osteoblast cell line, and surface cell proliferation was comparable with proliferation on reference polystyrene culture plates.


Assuntos
Durapatita/química , Teste de Materiais/métodos , Temperatura , Líquidos Corporais/química , Comunicação Celular/efeitos dos fármacos , Morte Celular/efeitos dos fármacos , Linhagem Celular , Sobrevivência Celular/efeitos dos fármacos , Materiais Revestidos Biocompatíveis/química , Materiais Revestidos Biocompatíveis/farmacologia , Cristalização , Análise Diferencial Térmica , Durapatita/farmacologia , Módulo de Elasticidade/efeitos dos fármacos , Dureza/efeitos dos fármacos , Humanos , Microscopia Eletrônica de Varredura , Gases em Plasma/química , Pós , Análise Espectral Raman , Eletricidade Estática , Suspensões , Resistência à Tração/efeitos dos fármacos , Termogravimetria , Difração de Raios X
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA