Acta Neurochir (Wien) ; 154(1): 33-41, 2012 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-22052473

RESUMO

OBJECTIVE: Epidemiological studies indicate a link between low-dose irradiation (<10,000 mGy) to the head and the local occurrence of tumors after decades of delay. Comparable radiation doses can be reached during neuro-endovascular procedures (NEP), but the incidence of similar exposures has not been completely delineated. We compared the levels of radiation to the head measured during NEP to those reported for patients developing radiation-induced cancers. METHODS: In our prospective study we determined the cumulative maximum entrance skin doses (MESD) and the incidence of epilation in 107 consecutive patients submitted to NEP between 2003 and 2007. We also extensively searched the literature and compared our results with the data we found. RESULTS: The cumulative MESD due to NEP was above 3,000 mGy (range 3,101-5,421 mGy) in 18 patients. In 22 we observed partial epilation within 10 weeks from the initial NEP. Sixty cases of epilation after NEP have been previously reported in the literature. The average of the reported MESD was 4,241 mGy (range 2,000-6,640 mGy). CONCLUSION: Physical dosimetry and the incidence of partial epilation indicate that about one fifth of the patients submitted to NEP received radiation doses comparable to those linked to the occurrence of tumors. The potential risks of developing tumors after a long delay, when compared to the immediate benefits of endovascular treatment of aneurysm and arteriovenous malformations (AVM) of the brain, do not counterindicate NEP, but increased awareness of the risk should help physicians and patients to make a fully informed decision when other treatments are available.


Assuntos
Angiografia Cerebral/efeitos adversos , Embolização Terapêutica/efeitos adversos , Procedimentos Endovasculares/efeitos adversos , Neoplasias Induzidas por Radiação/epidemiologia , Lesões por Radiação/epidemiologia , Tomografia Computadorizada por Raios X/efeitos adversos , Adolescente , Adulto , Idoso , Idoso de 80 Anos ou mais , Angiografia Cerebral/métodos , Criança , Embolização Terapêutica/métodos , Procedimentos Endovasculares/métodos , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Neoplasias Induzidas por Radiação/fisiopatologia , Estudos Prospectivos , Lesões por Radiação/fisiopatologia , Tomografia Computadorizada por Raios X/métodos , Adulto Jovem
3.
Cardiovasc Intervent Radiol ; 36(2): 320-9, 2013 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-23150118

RESUMO

PURPOSE: To measure the maximum entrance skin dose (MESD) on patients undergoing carotid artery stenting (CAS) using embolic-protection devices, to analyze the dependence of dose and exposure parameters on anatomical, clinical, and technical factors affecting the procedure complexity, to obtain some local diagnostic reference levels (DRLs), and to evaluate whether overcoming DRLs is related to procedure complexity. MATERIALS AND METHODS: MESD were evaluated with radiochromic films in 31 patients (mean age 72 ± 7 years). Five of 33 (15 %) procedures used proximal EPD, and 28 of 33 (85 %) procedures used distal EPD. Local DRLs were derived from the recorded exposure parameters in 93 patients (65 men and 28 women, mean age 73 ± 9 years) undergoing 96 CAS with proximal (33 %) or distal (67 %) EPD. Four bilateral lesions were included. RESULTS: MESD values (mean 0.96 ± 0.42 Gy) were <2 Gy without relevant dependence on procedure complexity. Local DRL values for kerma area product (KAP), fluoroscopy time (FT), and number of frames (NFR) were 269 Gy cm(2), 28 minutes, and 251, respectively. Only simultaneous bilateral treatment was associated with KAP (odds ratio [OR] 10.14, 95 % confidence interval [CI] 1-102.7, p < 0.05) and NFR overexposures (OR 10.8, 95 % CI 1.1-109.5, p < 0.05). Type I aortic arch decreased the risk of FT overexposure (OR 0.4, 95 % CI 0.1-0.9, p = 0.042), and stenosis ≥ 90 % increased the risk of NFR overexposure (OR 2.8, 95 % CI 1.1-7.4, p = 0.040). At multivariable analysis, stenosis ≥ 90 % (OR 2.8, 95 % CI 1.1-7.4, p = 0.040) and bilateral treatment (OR 10.8, 95 % CI 1.1-109.5, p = 0.027) were associated with overexposure for two or more parameters. CONCLUSION: Skin doses are not problematic in CAS with EPD because these procedures rarely lead to doses >2 Gy.


Assuntos
Estenose das Carótidas/cirurgia , Dispositivos de Proteção Embólica , Doses de Radiação , Radiografia Intervencionista , Radiometria/instrumentação , Pele/efeitos da radiação , Stents , Idoso , Idoso de 80 Anos ou mais , Angiografia , Feminino , Fluoroscopia , Humanos , Embolia Intracraniana/etiologia , Embolia Intracraniana/prevenção & controle , Modelos Lineares , Modelos Logísticos , Masculino , Pessoa de Meia-Idade , Resultado do Tratamento
4.
J Vasc Interv Radiol ; 17(5): 855-62, 2006 May.
Artigo em Inglês | MEDLINE | ID: mdl-16687752

RESUMO

PURPOSE: To investigate a method for evaluation of the maximum skin dose (MSD) of radiation in patients undergoing interventional radiology procedures in the liver. MATERIALS AND METHODS: Transarterial chemoembolization, transarterial embolization, portal vein embolization, and biliary interventions were the procedures considered in this study. Ninety procedures in 70 patients were studied. The complexity of the biliary interventions was taken into account during the analysis. The MSD was measured with use of GAF chromic XR type R films, whereas the dose-area product (DAP) was measured by a transmission chamber. RESULTS: The MSD was measured in a group of 19 patients. The coefficient of the interpolation line between the skin dose and the DAP (0.0051 cm(-2)) was determined. An approximated value of MSD from the DAP for the remaining 71 procedures was estimated by means of an interpolation line. The mean MSD in the endovascular procedures was 4.4 Gy; for the three different levels of complexity of the biliary procedures, the mean MSDs were 0.4, 1.2, and 3 Gy, respectively. CONCLUSIONS: Radiochromic films are an easy-to-use and efficient method for measuring skin entrance radiation dose and have the advantage of providing information on the MSD as well as the distribution of radiation to the skin. In light of these results, the authors suggest recording the dose for biliary and endovascular procedures as indicated by the International Commission on Radiological Protection and the United States Food and Drug Administration.


Assuntos
Doenças Biliares/diagnóstico por imagem , Quimioembolização Terapêutica/métodos , Hepatopatias/radioterapia , Monitoramento de Radiação/métodos , Radiografia Intervencionista , Adulto , Idoso , Idoso de 80 Anos ou mais , Relação Dose-Resposta à Radiação , Feminino , Fluoroscopia , Humanos , Masculino , Pessoa de Meia-Idade , Imagens de Fantasmas , Doses de Radiação , Monitoramento de Radiação/instrumentação