Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
J Am Soc Mass Spectrom ; 24(1): 109-14, 2013 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-23345059

RESUMO

As the resolution of analytical methods improves, further progress tends to be increasingly limited by instrumental parameter instabilities that were previously inconsequential. This is now the case with differential ion mobility spectrometry (FAIMS), where fluctuations of the voltages and gas pressure have become critical. A new high-definition generator for FAIMS compensation voltage reported here provides a stable and accurate output than can be scanned with negligible steps. This reduces the spectral drift and peak width, thus improving the resolving power (R) and resolution. The gain for multiply-charged peptides that have narrowest peaks is up to ~40%, and R ~400-500 is achievable using He/N(2) or H(2)/N(2) gas mixtures.


Assuntos
Espectrometria de Massas/instrumentação , Espectrometria de Massas/métodos , Gases/química , Isomerismo , Peptídeos/química
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA