1.
BMC Pediatr ; 15: 206, 2015 Dec 11.
Artigo em Inglês | MEDLINE | ID: mdl-26654316

RESUMO

BACKGROUND: Random safety audits (RSA) are a safety tool enabling prevention of adverse events, but they have not been widely used in hospitals. The aim of this study was to use RSAs to assess and compare the frequency of appropriate use of infusion pump safety systems in a Neonatal Intensive Care Unit (NICU) before and after quality improvement interventions and to analyse the intravenous medication programming data. METHODS: Prospective, observational study comparing the frequency of appropriate use of Alaris® CC smart pumps through RSAs over two periods, from 1 January to 31 December 2012 and from 1 November 2014 to 31 January 2015. Appropriate use was defined as all evaluated variables being correctly programmed into the same device. Between the two periods they were established interventions to improve the use of pumps. The information recorded at the pumps with the new security system, also extracted for one year. RESULTS: Fifty-two measurements were collected during the first period and 160 measurements during the second period. The frequency of appropriate use was 73.13 % (117/160) in the second period versus 0 % (0/52) in the first period (p < 0.0001). Information was recorded on 44,924 infusions; in 46.03 % (20,680/44,924) of cases the drug name was recorded. In 2.5 % (532/20,680) of cases there was an attempt to exceed the absolute limit. CONCLUSIONS: Random Safety Audits were a very useful tool for detecting inappropriate use of pumps in the NICU. The improvement strategies were effective for improving appropriate use and programming of the intravenous medication infusion pumps in our NICU.


Assuntos
Bombas de Infusão , Unidades de Terapia Intensiva Neonatal/normas , Auditoria Médica , Segurança do Paciente/normas , Humanos , Recém-Nascido , Sistemas de Medicação no Hospital/normas , Estudos Prospectivos , Melhoria de Qualidade , Gestão da Segurança
2.
Biomed Res Int ; 2015: 719497, 2015.
Artigo em Inglês | MEDLINE | ID: mdl-26558277

RESUMO

BACKGROUND: Random safety audits (RSAs) are a safety tool but have not been widely used in hospitals. OBJECTIVES: To determine the frequency of proper use of equipment safety mechanisms in relation to monitoring and mechanical ventilation by performing RSAs. The study also determined whether factors related to the patient, time period, or characteristics of the area of admission influenced how the device safety systems were used. METHODS: A prospective observational study was conducted in a level III-C Neonatal Intensive Care Unit (NICU) during 2012. 87 days were randomly selected. Appropriate overall use was defined when all evaluated variables were correctly programmed in the audited device. RESULTS: A total of 383 monitor and ventilator audits were performed. The Kappa coefficient of interobserver agreement was 0.93. The rate of appropriate overall use of the monitors and respiratory support equipment was 33.68%. Significant differences were found with improved usage during weekends, OR 1.85 (1.12-3.06, p = 0.01), and during the late shift (3 pm to 10 pm), OR 1.59 (1.03-2.4, p = 0.03). CONCLUSIONS: Equipment safety systems of monitors and ventilators are not properly used. To improve patient safety, we should identify which alarms are really needed and where the difficulties lie for the correct alarm programming.


Assuntos
Unidades de Terapia Intensiva Neonatal/normas , Monitorização Fisiológica/instrumentação , Monitorização Fisiológica/normas , Respiração Artificial/instrumentação , Respiração Artificial/normas , Auditoria Clínica , Humanos , Recém-Nascido , Segurança do Paciente , Estudos Prospectivos