Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Assunto principal
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Annu Int Conf IEEE Eng Med Biol Soc ; 2020: 5259-5262, 2020 07.
Artigo em Inglês | MEDLINE | ID: mdl-33019170

RESUMO

Gastric ablation is emerging as a potential therapy for electrical dysrhythmias associated with gastric disorders. Thermal conduction properties of gastric tissue during ablation have not yet been defined, but are necessary for optimizing the technique and translating ablation to clinical therapy. We developed custom needle-based transmural temperature probes to quantify the temperature of gastric tissue during ablation. These probes were applied in vivo in pigs (n=5), during gastric ablation (70 °C, 10 s duration), at distances of 2.5 - 20 mm from the ablation catheter tip. Thermal response of the tissue was non-linear; the maximum temperature increase from baseline (33.3 ± 1.0 °C) was observed at the closest temperature probe to the catheter tip (2.5 mm, 14.9 °C), and temperature change decreased with distance from the catheter tip. Probes positioned between 5 -20 mm from the catheter tip recorded temperature increases of less than 5.6 °C. This study provides methods for monitoring temperature during in vivo ablation, and demonstrates that functional temperature increases from ablation were restricted to within approximately 5 mm of the catheter. These methods can now be applied to optimize effective ablation parameters, and to inform models of gastric ablation.


Assuntos
Ablação por Cateter , Animais , Primeiros Socorros , Suínos , Temperatura , Condutividade Térmica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA