Dose calibrator linearity test: 99mTc versus 18F radioisotopes / Teste de linearidade em medidor de atividade: utilização do radioisótopo 99mTc versus 18F
Willegaignon, José; Sapienza, Marcelo Tatit; Coura-Filho, George Barberio; Garcez, Alexandre Teles; Alves, Carlos Eduardo Gonzalez Ribeiro; Cardona, Marissa Anabel Rivera; Gutterres, Ricardo Fraga; Buchpiguel, Carlos Alberto.
Afiliação
  • Willegaignon, José; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Sapienza, Marcelo Tatit; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Coura-Filho, George Barberio; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Garcez, Alexandre Teles; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Alves, Carlos Eduardo Gonzalez Ribeiro; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Cardona, Marissa Anabel Rivera; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Gutterres, Ricardo Fraga; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
  • Buchpiguel, Carlos Alberto; Instituto do Câncer do Estado de São Paulo Octavio Frias de Oliveira. Service of Nuclear Medicine. São Paulo. BR
Radiol. bras ; Radiol. bras;48(1): 26-32, Jan-Feb/2015. tab, graf
Article em En | LILACS | ID: lil-741690
Biblioteca responsável: BR1.1
ABSTRACT

Objective:

The present study was aimed at evaluating the viability of replacing 18F with 99mTc in dose calibrator linearity testing. Materials and

Methods:

The test was performed with sources of 99mTc (62 GBq) and 18F (12 GBq) whose activities were measured up to values lower than 1 MBq. Ratios and deviations between experimental and theoretical 99mTc and 18F sources activities were calculated and subsequently compared.

Results:

Mean deviations between experimental and theoretical 99mTc and 18F sources activities were 0.56 (± 1.79)% and 0.92 (± 1.19)%, respectively. The mean ratio between activities indicated by the device for the 99mTc source as measured with the equipment pre-calibrated to measure 99mTc and 18F was 3.42 (± 0.06), and for the 18F source this ratio was 3.39 (± 0.05), values considered constant over the measurement time.

Conclusion:

The results of the linearity test using 99mTc were compatible with those obtained with the 18F source, indicating the viability of utilizing both radioisotopes in dose calibrator linearity testing. Such information in association with the high potential of radiation exposure and costs involved in 18F acquisition suggest 99mTc as the element of choice to perform dose calibrator linearity tests in centers that use 18F, without any detriment to the procedure as well as to the quality of the nuclear medicine service. .
RESUMO

Objetivo:

Avaliar a viabilidade de substituir o radioisótopo 18F pelo 99mTc no teste de linearidade em medidores de atividade.

Materiais e Métodos:

O teste foi realizado empregando-se fontes de 99mTc (62 GBq) e 18F (12 GBq), cujas atividades foram mensuradas até valores abaixo de 1 MBq. As razões e desvios entre as atividades experimentais e teóricas foram calculados e comparados.

Resultados:

Os desvios médios entre as atividades experimentais e teóricas para o 99mTc e 18F foram, respectivamente, 0,56 (± 1,79)% e 0,92 (± 1,19)%. A razão média entre as atividades indicadas pelo equipamento para a fonte de 99mTc quando mensurada no equipamento pré-ajustado para medir o 99mTc e 18F foi 3,42 (± 0,06), e para o 18F este valor foi 3,39 (± 0,05), razões consideradas constantes ao longo de todo o período de medida.

Conclusão:

Os resultados do teste utilizando o 99mTc são compatíveis com os adquiridos com o 18F, implicando na possibilidade de utilização de ambos os radioisótopos na realização do teste de linearidade. Esta informação, aliada ao elevado potencial de exposição e custos de aquisição do 18F, sugere que o 99mTc seja empregado na realização do teste de linearidade para clínicas que utilizam 18F, sem prejuízo para o procedimento e garantia da qualidade de um serviço de medicina nuclear. .