Your browser doesn't support javascript.
loading
Comparison of phenyl-type columns in the development of a fast liquid chromatographic system for eighteen opiates commonly found in forensic toxicology.
Dams, R; Lambert, W E; Clauwaert, K M; De Leenheer, A P.
Afiliação
  • Dams R; Laboratorium voor Toxicologie, Universiteit Gent, Ghent, Belgium.
J Chromatogr A ; 896(1-2): 311-9, 2000 Oct 27.
Article em En | MEDLINE | ID: mdl-11093666
ABSTRACT
We report a precise and reliable method for the detection of 18 of the most commonly found opiates on the Belgian legal and illicit market, by ion-exchange, reversed-phase high-performance liquid chromatography, using a conventional phenyl-type analytical column (150x4.6 mm I.D., particle size 5 microm) and diode-array detection. We also describe a performance (efficiency and sensitivity) comparison of this column to a recently developed "high-speed" column (53x7.0 mm I.D., particle size 3 microm) packed with the same stationary phase, and used under slightly adjusted flow and gradient conditions. The final method, using the "high-speed" column, showed a significant reduction (55%) in analysis time without loss of resolution and sensitivity.