Your browser doesn't support javascript.
loading
Evaluation of 11-gauge and 9-gauge vacuum-assisted breast biopsy systems in a breast parenchymal model.
Poellinger, Alexander; Bick, Ulrich; Freund, Torsten; Diekmann, Susanne; Hamm, Bernd; Diekmann, Felix.
Afiliação
  • Poellinger A; Department of Radiology, Campus Charité Mitte, Schumannstr. 20/21, D-10098 Berlin, Germany. alexander.poellinger@charite.de
Acad Radiol ; 14(6): 677-84, 2007 Jun.
Article em En | MEDLINE | ID: mdl-17502257
ABSTRACT
RATIONALE AND

OBJECTIVES:

To compare three commercially available vacuum-assisted breast biopsy systems for tissue yield, length and fragmentation of specimens. MATERIALS AND

METHODS:

Specimens were acquired from radiolucent (bacon) and radioopaque (turkey breast) tissue using three different commercially available vacuum-assisted breast biopsy devices. Two systems (system 1 and 2) were equipped with 11 G needles, one system (system 3) with a 9 G needle. As for systems 1 and 2 a second chamber for applying the vacuum is attached to the needle, the external maximum diameter was identical for all three systems. 48 specimens were taken out for each tissue type and for each device. Specimens were measured for total weight, individual length, and number of fragments. Differences between groups were analyzed using analysis of variance (ANOVA) and Student's t-test.

RESULTS:

For both tissue types, system 1 and 2 showed similar results, for system 3 tissue weight and length of specimens were larger. Differences in lengths and weight were statistically significant between system 1 and 3 and system 2 and 3 (ANOVA, P < 0.05). Differences between length and weight were statistically significant between tissue 1 and 2 for all devices (t-Test < 0.05).

CONCLUSION:

As for system 3 a larger tissue yield was obtained with the same number of specimens compared to systems 1 and 2, it can be assumed that the same diagnostic accuracy as for systems 1 and 2 may be achieved for system 3 with less passes through tissue.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2007 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Animals Idioma: En Ano de publicação: 2007 Tipo de documento: Article