Absolute efficiency estimation of photon-number-resolving detectors using twin beams.
Worsley, A P; Coldenstrodt-Ronge, H B; Lundeen, J S; Mosley, P J; Smith, B J; Puentes, G; Thomas-Peter, N; Walmsley, I A.
Afiliação
  • Worsley AP; University of Oxford, Clarendon Laboratory, Parks Road, Oxford, OX1 3PU, United Kingdom.
Opt Express ; 17(6): 4397-411, 2009 Mar 16.
Article em En | MEDLINE | ID: mdl-19293867
ABSTRACT
A nonclassical light source is used to demonstrate experimentally the absolute efficiency calibration of a photon-number-resolving detector. The photon-pair detector calibration method developed by Klyshko for single-photon detectors is generalized to take advantage of the higher dynamic range and additional information provided by photon-number-resolving detectors. This enables the use of brighter twin-beam sources including amplified pulse pumped sources, which increases the relevant signal and provides measurement redundancy, making the calibration more robust.