Blocked impurity band detectors applied to tunable diode laser spectroscopy in the 8- to 28-microm range.
Appl Opt ; 32(12): 2117-21, 1993 Apr 20.
Article em En | MEDLINE | ID: mdl-20820356
ABSTRACT
A novel tunable diode laser spectrometer operating at 8 < lambda < 28 microm is described. A blocked impurity band SiAs chip is employed as detector. This device operates in this wavelength range with high detectivity and adequate frequency response for the high-sensitivity techniques used. A combination of sweep averaging and second-harmonic detection at 22 kHz yielded signal-to-noise ratios of 1200 at lambda > 20 microm. The sensitivity and spectral resolution achieved are 1 order of magnitude better than those of Fourier instruments in this range, with an improvement in instrument time response of ~ 3000. Several molecular bands of CO(2) and N(2)O are observed for what is, to our knowledge, the first time with this instrument. Examples of spectral line measurements are presented.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 1993 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 1993 Tipo de documento: Article