Swept source optical coherence tomography measurement of the iris-trabecular contact (ITC) index: a new parameter for angle closure.
Ho, Sue-Wei; Baskaran, Mani; Zheng, Ce; Tun, Tin A; Perera, Shamira A; Narayanaswamy, Arun K; Friedman, David S; Aung, Tin.
Afiliação
  • Ho SW; Singapore Eye Research Institute and Singapore National Eye Centre, Singapore, Singapore.
Graefes Arch Clin Exp Ophthalmol ; 251(4): 1205-11, 2013 Apr.
Article em En | MEDLINE | ID: mdl-23001586
ABSTRACT

PURPOSE:

To evaluate the inter- and intra-observer agreement of measurement of the iris-trabecular contact (ITC) index, a measure of the degree of angle closure, using swept source optical coherence tomography (SSOCT, CASIA SS-1000, Tomey Corporation, Nagoya, Japan).

METHODS:

One randomly selected eye of 60 subjects was imaged under dark room conditions. The SSOCT 3-dimensional angle scan simultaneously obtains 128 radial scans of the anterior chamber for the entire circumference of the angle. Post-imaging analysis estimated the ITC index using in-built software. For intra-observer agreement for image grading, one examiner performed the grading twice in a masked fashion and random order after a 1-week interval. A second examiner graded images to assess inter-observer agreement for image grading. For intra-observer agreement for image acquisition, a single operator imaged patients twice. For inter-observer agreement for image acquisition, a single observer graded two sets of images acquired by two different operators on the same patient. Bland-Altman plots and 95 % limits of agreement (LOA) were reported.

RESULTS:

Study subjects were predominantly Chinese (54/60, 90 %) and female (42/60, 70 %), with a mean age of 65.5 years. The median ITC index for eyes with open angles (31/60) and closed angles was 20 % (95 % confidence interval [CI] - 13.6, 27.8) and 49 % (95%CI - 35.5, 69.2) respectively. The mean difference (95 % LOA) for intra-observer agreement for image grading and image acquisition were -0.8 % (-8.2, 6.5) and 0.6 % (-10.9, 9.7); corresponding inter- observer agreement were 0.1 % (-10, 10.1) and -0.3 % (-11.1, 10.5) respectively.

CONCLUSIONS:

The inter- and intra-observer agreement of the ITC index, as a measure of extent of angle closure using SSOCT, was good.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies / Prognostic_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2013 Tipo de documento: Article