Your browser doesn't support javascript.
loading
[Measurement and analysis of lead in soil using X-ray fluorescence spectroscopy].
Zhang, Rong; Zhang, Yu-Jun; Zhang, Wei; Chen, Dong; Yu, Xiao-Ya; Gao, Yan-Wei.
Afiliação
  • Zhang R; Key Laboratory of Environment Optics and Technology, Chinese Academy of Sciences, Hefei 230031, China. rzhang@aiofm.ac.cn
Guang Pu Xue Yu Guang Pu Fen Xi ; 33(2): 554-7, 2013 Feb.
Article em Zh | MEDLINE | ID: mdl-23697153
The present paper analyzed the characteristics of X-ray fluorescence spectroscopy (XRF) of metal element lead in soil using the NITON XLt793 portable X-ray fluorescence spectra of heavy metal analyzer under laboratory conditions. The characteristic spectral lines of L(alpha) (energy: 10. 55 keV) and L(beta) (energy: 12. 61 keV) with different matrix elements were selected respectively for lead in the experiment. By measuring the intensities of the characteristic spectral line with different Pb concentration, the results demonstrate that the relation between concentration [mass fraction 10 x 10(-6) - 1 800 x 10(-6)] of Pb element and the intensity of the characteristic spectrum is well linear. The calibration curve of Pb was plotted based on the different concentration measurement results, and the limit of detection of 7.89 x 10(-6) was obtained for Pb in soil.
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Idioma: Zh Ano de publicação: 2013 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Idioma: Zh Ano de publicação: 2013 Tipo de documento: Article