Your browser doesn't support javascript.
loading
Treatment of PPROM with anhydramnion in humans: first experience with different amniotic fluid substitutes for continuous amnioinfusion through a subcutaneously implanted port system.
J Perinat Med ; 41(6): 657-63, 2013 Nov.
Article em En | MEDLINE | ID: mdl-23774012
ABSTRACT

OBJECTIVE:

This study aims to treat patients with preterm premature rupture of the membranes (PPROM) and anhydramnion using continuous amnioinfusion through a subcutaneously implanted port system.

METHODS:

An amniotic fluid replacement port system was implanted in seven patients with PPROM and anhydramnion starting at the 20th week of gestation (range, 14-26 weeks) for long-term amnioinfusion. Saline solutions (2 L/day; Jonosteril(®), Sterofundin(®), isotonic NaCl 0.9% solution, lactated Ringer's solution) and a hypotonic aqueous composition with reduced chloride content similar to the electrolyte concentration of human amniotic fluid were used for the continuous amnioinfusion.

RESULTS:

The mean duration of the PPROM delivery interval continued for 49 days (range, 9-69 days), with 3 weeks of amnioinfusion via the port system (range, 4-49). The newborns showed no signs of lung hypoplasia.

CONCLUSION:

Long-term lavage of the amniotic cavity via a subcutaneously implanted port system in patients with PPROM and anhydramnion may help prolong the pregnancy and avoid fetal lung hypoplasia. A hypotonic aqueous composition with reduced chloride content similar to human amniotic fluid can be safely used for amnioinfusion. Prospective randomized studies are ongoing.