Improvement and validation of d-xylose determination in urine and serum as a new tool for the noninvasive evaluation of lactase activity in humans.
Hermida, Carmen; Martínez-Costa, Oscar H; Corrales, Guillermo; Teruel, Cristina; Sánchez, Valentina; Sánchez, José J; Sarrión, Dolores; Ariza, María J; Codoceo, Rosa; Calvo, Idoya; Fernández-Mayoralas, Alfonso; Aragón, Juan J.
Afiliação
  • Hermida C; Departamento de Bioquímica and Instituto de Investigaciones Biomédicas Alberto Sols, Facultad de Medicina de la Universidad Autónoma de Madrid, Universidad Autónoma de Madrid-Consejo Superior de Investigaciones Científicas, Madrid, Spain; Venter Pharma SL, Madrid, Spain.
J Clin Lab Anal ; 28(6): 478-86, 2014 Nov.
Article em En | MEDLINE | ID: mdl-24659338
ABSTRACT

BACKGROUND:

The phloroglucinol assay is the current method for d-xylose determination in urine/plasma/serum. However, its sensitivity is limited when low amounts of d-xylose are to be measured, such as in the noninvasive evaluation of intestinal lactase with 4-galactosylxylose (gaxilose). An improved assay was therefore needed.

METHODS:

We developed and validated a modified version of the phloroglucinol-based assay for quantification of d-xylose in urine/serum samples. A method for gaxilose determination by gas chromatography (GC) was also optimized.

RESULTS:

Linearity ranged from 0.125 to 5.0 mg/l (5-200 mg/l in original sample). Accuracy at LOQ (0.125 mg/l) was 0.97/2.49% in spiked urine/serum; for other quality controls (QC), it was <1.27%. Intra- and interassay precision at LOQ were 6.02% and 6.45% for urine, and 8.86% and 10.00%, respectively, for serum; for other QC, precision was <2.15%. Linearity of gaxilose determination by GC was 3.90-195.17 for urine and 9.75-195.17 mg/l for serum with acceptable sensitivity and reproducibility. The method proved adequate for the d-xylose determination in healthy and hypolactasic subjects after oral administration of gaxilose.

CONCLUSIONS:

The modified method provides high sensitivity and robustness for d-xylose quantification in urine/serum for routine clinical use especially in the noninvasive diagnosis of intestinal lactase deficiency with the gaxilose test.