Accuracy of Working Length Determination using NovApex and Root-ZX Apex Locators: An in vitro Study.
Zand, Vahid; Rahimi, Saeed; Davoudi, Paria; Afshang, Abubakr.
Afiliação
  • Zand V; Department of Endodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, East Azerbaijan, Islamic Republic of Iran.
  • Rahimi S; Department of Endodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, East Azerbaijan, Islamic Republic of Iran.
  • Davoudi P; Department of Endodontics, Faculty of Dentistry, Tabriz University of Medical Sciences, East Azerbaijan, Islamic Republic of Iran, Phone: +989144045938, e-mail: davoudiparia1991@gmail.com.
  • Afshang A; Private Office, East Azerbaijan, Islamic Republic of Iran.
J Contemp Dent Pract ; 18(5): 383-385, 2017 May 01.
Article em En | MEDLINE | ID: mdl-28512277
ABSTRACT

INTRODUCTION:

This ex vivo study evaluated the accuracy of the Root-ZX electronic apex locator (EAL) (J. Morita, Tokyo, Japan) and the NovApex (Forum Technologies, Rishon Lezion, Israel) in determining the working length (WL) during endodontic treatment. MATERIALS AND

METHODS:

Forty extracted single-rooted human teeth were selected for this study. The actual WL was measured with visual technique by a size #15 k-file under magnification. Then, the canal lengths were measured electronically with both Root-ZX and NovApex apex locators within ±0.5 and ±1 mm. Mean percentage of data was analyzed between groups using paired t-test, with a statistically significant level of p < 0.05.

RESULTS:

The accuracy of NovApex apex locator was 85% within ±0.5 mm and 92.5% within ±1 mm. The accuracy of Root-ZX apex locator was 70% within ±0.5 mm and 97.5% within ±1 mm. There was no significant difference between the accuracy of the two EALs.

CONCLUSION:

Both the NovApex and Root-ZX EALs are useful for measuring the WL with high accuracy. CLINICAL

SIGNIFICANCE:

Given the importance of accurate WL determination in the success of endodontic treatments, the accuracy of different apex locators should be evaluated.