Physicochemical and microscopic characterization of implant-abutment joints.
Lopes, Patricia A; Carreiro, Adriana F P; Nascimento, Rubens M; Vahey, Brendan R; Henriques, Bruno; Souza, Júlio C M.
Afiliação
  • Lopes PA; Department of Dentistry, Federal University of Rio Grande do Norte, Natal, RN, Brazil.
  • Carreiro AFP; Department of Dentistry, Federal University of Rio Grande do Norte, Natal, RN, Brazil.
  • Nascimento RM; Department of Materials Engineering, Federal University of Rio Grande do Norte, Natal, RN, Brazil.
  • Vahey BR; The Herman Ostrow School of Dentistry, University of Southern California, Los Angeles, CA, USA.
  • Henriques B; Center for Microelectromechanical Systems (CMEMS-MINHO), University of Minho, Guimarães, Portugal.
  • Souza JCM; Center for Microelectromechanical Systems (CMEMS-MINHO), University of Minho, Guimarães, Portugal.
Eur J Dent ; 12(1): 100-104, 2018.
Article em En | MEDLINE | ID: mdl-29657532
ABSTRACT

OBJECTIVE:

The purpose of this study was to investigate Morse taper implant-abutment joints by chemical, mechanical, and microscopic analysis. MATERIALS AND

METHODS:

Surfaces of 10 Morse taper implants and the correlated abutments were inspected by field emission gun-scanning electron microscopy (FEG-SEM) before connection. The implant-abutment connections were tightened at 32 Ncm. For microgap evaluation by FEG-SEM, the systems were embedded in epoxy resin and cross-sectioned at a perpendicular plane of the implant-abutment joint. Furthermore, nanoindentation tests and chemical analysis were performed at the implant-abutment joints. STATISTICS Results were statistically analyzed via one-way analysis of variance, with a significance level of P < 0.05.

RESULTS:

Defects were noticed on different areas of the abutment surfaces. The minimum and maximum size of microgaps ranged from 0.5 µm up to 5.6 µm. Furthermore, defects were detected throughout the implant-abutment joint that can, ultimately, affect the microgap size after connection. Nanoindentation tests revealed a higher hardness (4.2 ± 0.4 GPa) for abutment composed of Ti6Al4V alloy when compared to implant composed of commercially pure Grade 4 titanium (3.2 ± 0.4 GPa).

CONCLUSIONS:

Surface defects produced during the machining of both implants and abutments can increase the size of microgaps and promote a misfit of implant-abutment joints. In addition, the mismatch in mechanical properties between abutment and implant can promote the wear of surfaces, affecting the size of microgaps and consequently the performance of the joints during mastication.