Your browser doesn't support javascript.
loading
Three-harmonic optimal multisine input power spectrum for bioimpedance identification.
Kwon, H; Rojas, C R; Rutkove, S B; Sanchez, B.
Afiliação
  • Kwon H; College of Science & Technology, Yonsei University, Wonju 26493, Republic of Korea.
Physiol Meas ; 40(5): 05NT02, 2019 06 04.
Article em En | MEDLINE | ID: mdl-31026854
OBJECTIVE: The attention towards optimal signals for measuring electrical bioimpedance (EBI) has increased recently due to the many advantages they offer such as, for example, reduced measuring time. Here, the design of a three-harmonic optimized multisine input power spectrum for measuring bioimpedance is considered. APPROACH: The approach is based on designing the input power spectrum multisine excitation by optimizing a scalar function of the information matrix using the Fricke-Morse model. MAIN RESULTS: Simple analytical equations are provided that can be adopted by the reader to optimize the multisine input power spectrum to satisfy any specific application-related EBI measurement. SIGNIFICANCE: The presented signal may be a good choice of excitation signal in EBI applications where optimal excitation power spectrum is desired.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Ano de publicação: 2019 Tipo de documento: Article