Inter-fractional variation of markers and applicators in single-implant high-dose-rate interstitial brachytherapy for gynecologic malignancies.
Ostyn, Mark; Burke, Aidan M; Fields, Emma; Todor, Dorin.
Afiliação
  • Ostyn M; Virginia Commonwealth University Health System, Department of Radiation Oncology, 401 College St., Richmond, VA 23298.
  • Burke AM; Georgetown University, 3700 O St NW, Washington, DC 20057.
  • Fields E; Virginia Commonwealth University Health System, Department of Radiation Oncology, 401 College St., Richmond, VA 23298.
  • Todor D; Virginia Commonwealth University Health System, Department of Radiation Oncology, 401 College St., Richmond, VA 23298. Electronic address: dorin.todor@vcuhealth.org.
Brachytherapy ; 20(4): 771-780, 2021.
Article em En | MEDLINE | ID: mdl-33994342
ABSTRACT

PURPOSE:

Implanted fiducial markers are a commonly used tool in delineating the CTV in high-dose-rate interstitial brachytherapy (HDR-ISBT) for gynecologic malignancy, but their reliability in gynacological sites is not well understood. These markers and interstitial applicators can experience interfractional motion due to organ swelling or other anatomical changes. The purpose of this study was to evaluate the spatial variation of these features. METHODS AND MATERIALS The spatial positions of 50 implanted markers and 202 needles were tracked in 15 patients treated over 70 fractions of HDR brachytherapy. Marker and/or needle coordinates were extracted from CT images with contours and dose distributions. Automated analysis determined marker self-consistency and displacements between various elements of the implant.

RESULTS:

From start to end fraction, the relative positions of the markers experienced an average magnitude displacement of 4.5 ± 3.0 mm while the average displacement of the applicator tips was 11 ± 8 mm, relative to their respective centers of mass (CM).

CONCLUSIONS:

Markers implanted lateral and superior to the CTV experience greater drift than other implant locations.