Impact of chronic outward force on arterial responses of proximal and distal of long superficial femoral artery stent.
Li, Hu; Rha, Seung-Woon; Choi, Byoung Geol; Choi, Se Yeon; Moon, Sang Ki; Jang, Won Young; Kim, Woohyeun; Ahn, Ji Hun; Park, Sang-Ho; Choi, Woong Gil; Yang, Rui Feng; Bai, Wen Wei; Choi, Cheol Ung; Ryu, Yang Gi; Baek, Man Jong; Oh, Dong Joo.
Afiliação
  • Li H; Department of Medicine, The Second Affiliated Hospital of Kunming Medical University, Kunming, Yunnan, China.
  • Rha SW; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Choi BG; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea. swrha617@163.com.
  • Choi SY; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Moon SK; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Jang WY; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Kim W; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Ahn JH; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Park SH; Department of Cardiology, Soon Chun Hyang University Gumi Hospital, Gumi-si, Republic of Korea.
  • Choi WG; Department of Cardiology, Soonchunhyang University Cheonan Hospital Korea, Cheonan-Ii, Republic of Korea.
  • Yang RF; Department of Internal Medicine, School of Medicine, Konkuk University, Chungju, Republic of Korea.
  • Bai WW; Department of Medicine, The Second Affiliated Hospital of Kunming Medical University, Kunming, Yunnan, China.
  • Choi CU; Department of Medicine, The Second Affiliated Hospital of Kunming Medical University, Kunming, Yunnan, China.
  • Ryu YG; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Baek MJ; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
  • Oh DJ; Cardiovascular Center, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul, 08308, Republic of Korea.
BMC Cardiovasc Disord ; 21(1): 323, 2021 06 30.
Article em En | MEDLINE | ID: mdl-34193057
ABSTRACT

BACKGROUND:

Self-expanding nitinol stent (SENS) implantation is commonly oversized in the superficial femoral artery (SFA), and leads to chronic outward force (COF) and in-stent restenosis (ISR). This study aimed to investigate the impact of COF of oversizing SENS on ISR of SFA.

METHODS:

In patients with implanted SENS in SFA, intimal hyperplasia especially between proximal segment and distal segment was evaluated by quantitative angiography, and the impact of COF on mid-term angiographic outcomes was investigated. In addition, porcine model with implanted SENS was used to evaluate the impact of COF on angiographic and histopathologic outcomes at 1 month. Excised stented arteries were evaluated by histopathologic analysis.

RESULTS:

We analyzed 65 SENS in 61 patients with follow-up angiography at 6 months to 1 year. The baseline diameter was 6.8 ± 0.71 mm and length were 97.0 ± 33.8 mm for the SENS. The ratio of the diameter of the stent to the reference vessel was 1.3 ± 0.24 at the proximal portion and 1.53 ± 0.27 at the distal portion (P < 0.001). In the long SFA stent, stent-to-vessel ratio was significantly higher in the distal stent than in the proximal stent (1.3 ± 0.2 vs. 1.55 ± 0.25, P = 0.001). ISR incidence was higher at the distal stent (37.3% vs 52.6%, P = 0.029). All 11 pigs survived for 4 weeks after SENS implantation. The vessel diameter was 4.04 ± 0.40 mm (control group) vs 4.45 ± 0.63 mm (oversized group), and the implanted stent diameter was 5.27 ± 0.46 mm vs. 7.18 ± 0.4 mm (P = 0.001). The stent-to-vessel diameter ratio was 1.31 ± 0.12 versus 1.63 ± 0.20 (P < 0.001). After 4 weeks, restenosis % was 29.5 ± 12.9% versus 46.8 ± 21.5% (P = 0.016). The neointimal area was 5.37 ± 1.15 mm2 vs. 8.53 ± 5.18 mm2 (P = 0.05). The restenosis % was 39.34 ± 8.53% versus 63.97 ± 17.1% (P = 0.001).

CONCLUSIONS:

COF is an important cause of restenosis in the distal portion of the SFA stent. Optimal sizing of the SFA stent is important to reduce the incidence of restenosis. Therefore, COF was an important factor of restenosis following distal SFA stenting.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Animals / Female / Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article