Angioscopic Evaluation of Vascular Response After Fluoropolymer-Based Drug-Eluting Stent Implantation for Femoropopliteal Artery Lesions.
Circ J
; 87(3): 432-429, 2023 02 24.
Article
em En
| MEDLINE
| ID: mdl-36624062
ABSTRACT
BACKGROUND:
Although favorable clinical outcomes have been demonstrated for fluoropolymer-based paclitaxel-eluting stents (FP-DES) in the treatment of femoropopliteal lesions, the vascular response after implantation has not been systematically studied through intravascular imaging.MethodsâandâResults:
We angioscopically compared FP-DES 24 in the early phase (mean [±SD] 3±1 months), 26 in the middle phase (12±3 months), and 20 in the late phase (≥18 months) after implantation. The dominant neointimal coverage grade, heterogeneity of neointimal coverage grade, and thrombus adhesion in the stent segment were evaluated. Neointimal coverage was graded as follows Grade 0, stent struts exposed; Grade 1, struts bulging into the lumen, although covered; Grade 2, struts embedded in the neointima, but visible; Grade 3, struts fully embedded and invisible. Dominant neointimal coverage and heterogeneity grades were significantly higher in the middle and late phases than in the early phase (all P<0.05), but did not differ significantly between the middle and late phases. The incidence of thrombus adhesion was recorded for all stents in each of the 3 different phases.CONCLUSIONS:
The middle and late phases after FP-DES implantation were associated with significantly higher dominant neointimal coverage and heterogeneity grades than the early phase. However, thrombus adhesion was observed in all phases after FP-DES implantation. Arterial healing may not be completed even in the late phase after FP-DES implantation.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Limite:
Humans
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article