Significant Reduction of Radiation Exposure Using Specific Settings of the O-Arm for Percutaneous Cementoplasty in Accordance With the ALARA Principle.
Prod'homme, Marc; Grasset, Didier; Akiki, Alain; Helfer, Lionel; Dietrich, Gilles; Müller, Jonas Pierre; Tomasoni, Mattia; Sellami, Mahmoud; Boscherini, Duccio.
Afiliação
  • Prod'homme M; Clinic La Source, Neuro Orthopedic Center, Lausanne, Switzerland marcprod86@gmail.com.
  • Grasset D; Orthopedic Surgery Department, Riviera-Chablais Hospital, Rennaz, Switzerland.
  • Akiki A; Clinic La Source, Neuro Orthopedic Center, Lausanne, Switzerland.
  • Helfer L; Orthopedic Surgery Department, Riviera-Chablais Hospital, Rennaz, Switzerland.
  • Dietrich G; Orthopedic Surgery Department, Riviera-Chablais Hospital, Rennaz, Switzerland.
  • Müller JP; Orthopedic Surgery Department, Riviera-Chablais Hospital, Rennaz, Switzerland.
  • Tomasoni M; Orthopedic Surgery Department, Lausanne University Hospital, Lausanne, Switzerland.
  • Sellami M; Orthopedic Surgery Department, Riviera-Chablais Hospital, Rennaz, Switzerland.
  • Boscherini D; Ophtalmology Department, Jules-Gonin University Eye Hospital, University of Lausanne, Lausanne, Switzerland.
Int J Spine Surg ; 17(4): 534-541, 2023 Aug.
Article em En | MEDLINE | ID: mdl-37253626
ABSTRACT

BACKGROUND:

Three-dimensional (3D) navigation has become routinely used in spine surgery, allowing more accurate and safe procedures. However, radiation exposure related to the use of imaging is an unresolved issue, and information about it is relatively scarce. The "as low as reasonably achievable" (ALARA) principle aims to reduce the radiation exposure for the patients as low as possible. The objective of this study was to compare the effective dose related to the use of the O-arm in standard settings with adapted features for dose reduction during percutaneous cementoplasty.

METHODS:

From March 2021 to October 2022, all consecutive patients who underwent navigated percutaneous cementoplasty with the use of the O-arm were prospectively included. Demographic, operative, irradiation, and radiological data were collected. The main outcome was the effective dose (E) in millisievert (mSv). Secondary outcomes were the absolute risk of cancer (AR) in percent equivalent to a whole-body exposition, operative time, and radiological results according to Garnier. In group A, patients were operated on with standard settings of the O-arm, whereas in group B, navigation on the field of view, collimation, and low-dose settings were used.

RESULTS:

A total of 70 patients were included in the study 43 in group A and 27 in group B. Also, 109 vertebrae were operated 59 in group A and 50 in group B. Mean E was significantly higher in group A than in group B (9.94 and 4.34 mSv, respectively; P < 0.01). The 3D-related E followed the same trend (7.82 and 3.97 mSv, respectively), as did 2-dimensional-related E (2.12 and 0.37 mSv, respectively; P < 0.01). Average AR was also significantly higher in group A than in group B (5.10-4% and 2.10-4% respectively; P < 0.01). Operative time was similar in both groups, but the rate of satisfactory radiological results was higher in group A than in group B (95% and 84%, respectively; P = 0.11), and we found similar rates of cement leakage (22% and 24%, respectively; P = 0.71).

CONCLUSIONS:

The application of settings of the O-arm in accordance with the ALARA principle helped to significantly reduce the radiation exposure and should be routinely used for O-arm-assisted cementoplasty procedures. CLINICAL RELEVANCE This study details technical aspects and settings that may help users of the O-arm to decrease radiation exposure to patients and surgeons alike, especially in cementoplasty procedures, as well as in other procedures performed under O-arm guidance.