Characterization and commissioning of a new collaborative multi-modality radiotherapy platform.
Wang, Zhongfei; Sun, Xiaohuan; Wang, Wei; Zhang, Te; Chen, Liting; Duan, Jie; Feng, Siqi; Chen, Yinzhu; Wei, Zhiwei; Zang, Jian; Xiao, Feng; Zhao, Lina.
Afiliação
  • Wang Z; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Sun X; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Wang W; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Zhang T; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Chen L; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Duan J; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Feng S; Our United Corporation, 710018, Xi'an, Shaanxi Province, P.R. China.
  • Chen Y; Our United Corporation, 710018, Xi'an, Shaanxi Province, P.R. China.
  • Wei Z; Our United Corporation, 710018, Xi'an, Shaanxi Province, P.R. China.
  • Zang J; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China.
  • Xiao F; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China. xiao_frank@163.com.
  • Zhao L; Department of Radiation Oncology, Xijing Hospital, Fourth Military Medical University, 710032, Xi'an, Shaanxi Province, P.R. China. zhaolina@fmmu.edu.cn.
Phys Eng Sci Med ; 46(3): 981-994, 2023 Sep.
Article em En | MEDLINE | ID: mdl-37378823
TaiChi, a new multi-modality radiotherapy platform that integrates a linear accelerator, a focusing gamma system, and a kV imaging system within an enclosed O-ring gantry, was introduced into clinical application. This work aims to assess the technological characteristics and commissioning results of the TaiChi platform. The acceptance testing and commissioning were performed following the manufacturer's customer acceptance tests (CAT) and several AAPM Task Group (TG) reports/guidelines. Regarding the linear accelerator (linac), all applicable validation measurements recommended by the MPPG 5.a (basic photon beam model validation, intensity-modulated radiotherapy (IMRT)/volumetric-modulated arc therapy (VMAT) validation, end-to-end(E2E) tests, and patient-specific quality assurance (QA)) were performed. For the focusing gamma system, the absorbed doses were measured using a PTW31014 ion chamber (IC) and PTW60016 diode detector. EBT3 films and a PTW60016 diode detector were employed to measure the relative output factors (ROFs). The E2E tests were performed using PTW31014 IC and EBT3 films. The coincidences between the imaging isocenter and the linac/gamma mechanical isocenter were investigated using EBT3 films. The image quality was evaluated regarding the contrast-to-noise ratio (CNR), spatial resolution, and uniformity. All tests included in the CAT met the manufacturer's specifications. All MPPG 5.a measurements complied with the tolerances. The confidence limits for IMRT/VMAT point dose and dose distribution measurements were achieved according to TG-119. The point dose differences were below 1.68% and gamma passing rates (3%/2 mm) were above 95.1% for the linac E2E tests. All plans of patient-specific QA had point dose differences below 1.79% and gamma passing rates above 96.1% using the 3%/2 mm criterion suggested by TG-218. For the focusing gamma system, the differences between the calculated and measured absorbed doses were below 1.86%. The ROFs calculated by the TPS were independently confirmed within 2% using EBT3 films and a PTW60016 detector. The point dose differences were below 2.57% and gamma passing rates were above 95.3% using the 2%/1 mm criterion for the E2E tests. The coincidences between the imaging isocenter and the linac/gamma mechanical isocenter were within 0.5 mm. The image quality parameters fully complied with the manufacturer's specifications regarding the CNR, spatial resolution, and uniformity. The multi-modality radiotherapy platform complies with the CAT and AAPM commissioning criteria. The commissioning results demonstrate that this platform performs well in mechanical and dosimetry accuracy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline / Prognostic_studies Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Guideline / Prognostic_studies Idioma: En Ano de publicação: 2023 Tipo de documento: Article