Electric Field-Based Spatial Analysis of Noncontact Unipolar Electrograms to Map Regional Activation-Repolarization Intervals.
JACC Clin Electrophysiol
; 9(8 Pt 1): 1217-1231, 2023 08.
Article
em En
| MEDLINE
| ID: mdl-37558285
ABSTRACT
BACKGROUND:
Spatial heterogeneity in repolarization plays an important role in generating and sustaining cardiac arrhythmias. Reliable determination of repolarization times remains challenging.OBJECTIVES:
The goal of this study was to improve processing of densely sampled noncontact unipolar electrograms to yield reliable high-resolution activation and repolarization maps.METHODS:
Endocardial noncontact unipolar electrograms were both simulated and recorded in pig left ventricle. Electrical activity on the endocardial surface was processed in terms of a pseudo-electric field. Activation and repolarization times were calculated by using an amplitude-weighted average on QRS and T waves (ie, the E-field method). This was compared vs the conventional Wyatt method on unipolar electrograms. Timing maps were validated against timing on endocardial action potentials in a simulation study. In vivo, activation and repolarization times determined by using this alternative E-field method were validated against simultaneously recorded endocardial monophasic action potentials (MAPs).RESULTS:
Simulation showed that the E-field method provides viable measurements of local endocardial action potential activation and repolarization times. In vivo, correlation of E-field activation times with MAP activation times (rE = 0.76; P < 0.001) was similar to those of Wyatt (rWyatt = 0.80, P < 0.001; P[h1rE > rWyatt] = 0.82); for repolarization times, correlation improved significantly (rE = 0.96, P < 0.001; rWyatt = 0.82, P < 0.001; P[h1rE > rWyatt] < 0.00001). This resulted in improved correlations of activation-repolarization intervals to endocardial action potential duration on MAP (rE = 0.96, P < 0.001; rWyatt = 0.86, P < 0.001; P[h1rE > rWyatt] < 0.00001). Spatial beat-to-beat variation of repolarization could only be calculated by using the E-field methodology and correlated well with the MAP beat-to-beat variation of repolarization (rE = 0.76; P = 0.001).CONCLUSIONS:
The E-field method substantially enhances information from endocardial noncontact electrogram data, allowing for dense maps of activation and repolarization times and derived parameters.Palavras-chave