Your browser doesn't support javascript.
loading
Framework for Radiation Oncology Department-wide Evaluation and Implementation of Commercial Artificial Intelligence Autocontouring.
Maes, Dominic; Gates, Evan D H; Meyer, Juergen; Kang, John; Nguyen, Bao-Ngoc Thi; Lavilla, Myra; Melancon, Dustin; Weg, Emily S; Tseng, Yolanda D; Lim, Andrew; Bowen, Stephen R.
Afiliação
  • Maes D; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington; Department of Radiation Oncology, University of Washington, Seattle, Washington. Electronic address: Dominic.Maes@FredHutch.org.
  • Gates EDH; Department of Radiation Oncology, University of Washington, Seattle, Washington.
  • Meyer J; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington; Department of Radiation Oncology, University of Washington, Seattle, Washington.
  • Kang J; Department of Radiation Oncology, University of Washington, Seattle, Washington.
  • Nguyen BT; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington.
  • Lavilla M; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington.
  • Melancon D; Department of Radiation Oncology, University of Washington, Seattle, Washington.
  • Weg ES; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington; Department of Radiation Oncology, University of Washington, Seattle, Washington.
  • Tseng YD; Department of Radiation Oncology, University of Washington, Seattle, Washington; Clinical Research Division, Fred Hutchinson Cancer Center, Seattle, Washington.
  • Lim A; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington; Department of Radiation Oncology, University of Washington, Seattle, Washington; Department of Radiation Oncology, University of Southern California, Los Angeles, California.
  • Bowen SR; Department of Radiation Oncology, Fred Hutchinson Cancer Center, Seattle, Washington; Department of Radiation Oncology, University of Washington, Seattle, Washington; Department of Radiology, University of Washington, Seattle, Washington.
Pract Radiat Oncol ; 14(2): e150-e158, 2024.
Article em En | MEDLINE | ID: mdl-37935308
ABSTRACT

PURPOSE:

Artificial intelligence (AI)-based autocontouring in radiation oncology has potential benefits such as standardization and time savings. However, commercial AI solutions require careful evaluation before clinical integration. We developed a multidimensional evaluation method to test pretrained AI-based automated contouring solutions across a network of clinics. METHODS AND MATERIALS Curated data included 121 patient planning computed tomography (CT) scans with a total of 859 clinically approved contours used for treatment from 4 clinics. Regions of interest (ROIs) were generated with 3 commercial AI-based automated contouring software solutions (AI1, AI2, AI3) spanning the following disease sites brain, head and neck (H&N), thorax, abdomen, and pelvis. Quantitative agreement between AI-generated and clinical contours was measured by Dice similarity coefficient (DSC) and Hausdorff distance (HD). Qualitative assessment was performed by multiple experts scoring blinded AI-contours using a Likert scale. Workflow and usability surveying was also conducted.

RESULTS:

AI1, AI2, and AI3 contours had high quantitative agreement in 27.8%, 32.8%, and 34.1% of cases (DSC >0.9), performing well in pelvis (median DSC = 0.86/0.88/0.91) and thorax (median DSC = 0.91/0.89/0.91). All 3 solutions had low quantitative agreement in 7.4%, 8.8%, and 6.1% of cases (DSC <0.5), performing worse in brain (median DSC = 0.65/0.78/0.75) and H&N (median DSC = 0.76/0.80/0.81). Qualitatively, AI1 and AI2 contours were acceptable (rated 1-2) with at most minor edits in 70.7% and 74.6% of ROIs (2906 ratings), higher for abdomen (AI1 79.2%) and thorax (AI2 90.2%), and lower for H&N (29.0/35.6%). An end-user survey showed strong user preference for full automation and mixed preferences for accuracy versus total number of structures generated.

CONCLUSIONS:

Our evaluation method provided a comprehensive analysis of both quantitative and qualitative measures of commercially available pretrained AI autocontouring algorithms. The evaluation framework served as a roadmap for clinical integration that aligned with user workflow preference.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article