Your browser doesn't support javascript.
loading
Joint angle estimation during shoulder abduction exercise using contactless technology.
Barzegar Khanghah, Ali; Fernie, Geoff; Roshan Fekr, Atena.
Afiliação
  • Barzegar Khanghah A; KITE Research Institute, Toronto Rehabilitation Institute, University Health Network, 550 University Ave, Toronto, M5G 2A2, ON, Canada. ali.barzegarkhanghah@mail.utoronto.ca.
  • Fernie G; Institute of Biomedical Engineering, University of Toronto, 164 College St., Toronto, M5S 3G9, ON, Canada. ali.barzegarkhanghah@mail.utoronto.ca.
  • Roshan Fekr A; KITE Research Institute, Toronto Rehabilitation Institute, University Health Network, 550 University Ave, Toronto, M5G 2A2, ON, Canada.
Biomed Eng Online ; 23(1): 11, 2024 Jan 28.
Article em En | MEDLINE | ID: mdl-38281988
ABSTRACT

BACKGROUND:

Tele-rehabilitation, also known as tele-rehab, uses communication technologies to provide rehabilitation services from a distance. The COVID-19 pandemic has highlighted the importance of tele-rehab, where the in-person visits declined and the demand for remote healthcare rises. Tele-rehab offers enhanced accessibility, convenience, cost-effectiveness, flexibility, care quality, continuity, and communication. However, the current systems are often not able to perform a comprehensive movement analysis. To address this, we propose and validate a novel approach using depth technology and skeleton tracking algorithms.

METHODS:

Our data involved 14 participants (8 females, 6 males) performing shoulder abduction exercises. We collected depth videos from an LiDAR camera and motion data from a Motion Capture (Mocap) system as our ground truth. The data were collected at distances of 2 m, 2.5 m, and 3.5 m from the LiDAR sensor for both arms. Our innovative approach integrates LiDAR with the Cubemos and Mediapipe skeleton tracking frameworks, enabling the assessment of 3D joint angles. We validated the system by comparing the estimated joint angles versus Mocap outputs. Personalized calibration was applied using various regression models to enhance the accuracy of the joint angle calculations.

RESULTS:

The Cubemos skeleton tracking system outperformed Mediapipe in joint angle estimation with higher accuracy and fewer errors. The proposed system showed a strong correlation with Mocap results, although some deviations were present due to noise. Precision decreased as the distance from the camera increased. Calibration significantly improved performance. Linear regression models consistently outperformed nonlinear models, especially at shorter distances.

CONCLUSION:

This study showcases the potential of a marker-less system, to proficiently track body joints and upper-limb angles. Signals from the proposed system and the Mocap system exhibited robust correlation, with Mean Absolute Errors (MAEs) consistently below [Formula see text]. LiDAR's depth feature enabled accurate computation of in-depth angles beyond the reach of traditional RGB cameras. Altogether, this emphasizes the depth-based system's potential for precise joint tracking and angle calculation in tele-rehab applications.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Limite: Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Limite: Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article