Your browser doesn't support javascript.
loading
Metal artifact reduction sequence-MRI abnormalities in asymptomatic patients with dual-mobility hip prostheses.
Heckmann, Nathanael D; Chung, Brian C; Liu, Kevin C; Chen, Xiao T; Lovro, Luke R; Kistler, Natalie M; White, Eric; Christ, Alexander B; Longjohn, Donald B; Oakes, Daniel A; Lieberman, Jay R.
Afiliação
  • Heckmann ND; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Chung BC; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Liu KC; Keck School of Medicine of the University of Southern California, Los Angeles, California, USA.
  • Chen XT; Mayo Clinic, Rochester, Minnesota, USA.
  • Lovro LR; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Kistler NM; Keck School of Medicine of the University of Southern California, Los Angeles, California, USA.
  • White E; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Christ AB; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Longjohn DB; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Oakes DA; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
  • Lieberman JR; Department of Orthopaedic Surgery, Keck Medical Center of the University of Southern California, Los Angeles, California, USA.
Bone Joint J ; 106-B(3 Supple A): 89-96, 2024 Mar 01.
Article em En | MEDLINE | ID: mdl-38423076
ABSTRACT

Aims:

Modular dual-mobility (DM) articulations are increasingly used during total hip arthroplasty (THA). However, concerns remain regarding the metal liner modularity. This study aims to correlate metal artifact reduction sequence (MARS)-MRI abnormalities with serum metal ion levels in patients with DM articulations.

Methods:

A total of 45 patients (50 hips) with a modular DM articulation were included with mean follow-up of 3.7 years (SD 1.2). Enrolled patients with an asymptomatic, primary THA and DM articulation with over two years' follow-up underwent MARS-MRI. Each patient had serum cobalt, chromium, and titanium levels drawn. Patient satisfaction, Oxford Hip Score, and Forgotten Joint Score-12 (FJS-12) were collected. Each MARS-MRI was independently reviewed by fellowship-trained musculoskeletal radiologists blinded to serum ion levels.

Results:

Overall, two patients (4.4%) had abnormal periprosthetic fluid collections on MARS-MRI with cobalt levels > 3.0 µg/l. Four patients (8.9%) had MARS-MRI findings consistent with greater trochanteric bursitis, all with cobalt levels < 1.0 µg/l. A seventh patient had a periprosthetic fluid collection with normal ion levels. Of the 38 patients without MARS-MRI abnormalities, 37 (97.4%) had cobalt levels < 1.0 µg/l, while one (2.6%) had a cobalt level of 1.4 µg/l. One patient (2.2%) had a chromium level > 3.0 µg/l and a periprosthetic fluid collection. Of the 41 patients with titanium levels, five (12.2%) had titanium levels > 5.0 µg/l without associated MARS-MRI abnormalities.

Conclusion:

Periprosthetic fluid collections associated with elevated serum cobalt levels in patients with asymptomatic DM articulations occur infrequently (4.4%), but further assessment is necessary due to implant heterogeneity.