Delayed reaction to the Dacron buttress used in urethropexy.
J Reprod Med
; 29(9): 689-92, 1984 Sep.
Article
em En
| MEDLINE
| ID: mdl-6238164
ABSTRACT
Eleven patients developed a delayed reaction to the Dacron buttress used in Stamey urethropexy. The reaction was characterized by pain, dyspareunia, vaginal discharge and/or bleeding, induration of the abdominal incision, vaginal granulation tissue and draining abdominal or vaginal sinus tracts. This complication occurred in 5% of the 163 patients we treated with the Stamey procedure for stress urinary incontinence. Treatment consisted of removal of the nylon suture and Dacron buttress. None of the 11 patients developed recurrent incontinence. The subjective cure rate for stress urinary incontinence was 92% using the Stamey procedure.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Tipo de estudo:
Etiology_studies
Limite:
Adult
/
Aged
/
Female
/
Humans
/
Middle aged
Idioma:
En
Ano de publicação:
1984
Tipo de documento:
Article