Validation of volume flow measurements with cine phase-contrast MR imaging for peripheral arterial waveforms.
McCauley, T R; Peña, C S; Holland, C K; Price, T B; Gore, J C.
Afiliação
  • McCauley TR; Department of Diagnostic Radiology, Yale University School of Medicine, New Haven, CT, USA.
J Magn Reson Imaging ; 5(6): 663-8, 1995.
Article em En | MEDLINE | ID: mdl-8748483
ABSTRACT
A flow phantom was used to study MR volume flow measurements for monophasic and triphasic waveforms over the flow range expected in peripheral arteries at rest and with exercise (2-24 mL/sec, n = 50). The improvement in accuracy with phase-correction image processing to eliminate errors caused by eddy currents was measured. Volume flow estimates with Doppler sonography were also measured. MR volume flow measurements correlated with volume collection with r = .996 and mean error = 4.6%. Phase-correction processing decreased mean error from 12.6% to 4.6% (P < .001, paired t-test). Doppler sonography had a higher mean error of 10.3% (P < .001, unpaired t-test). Cine phase-contrast MR imaging provides accurate estimates of volume blood flow for waveforms and flow ranges expected in peripheral arteries.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Limite: Adult / Humans / Male Idioma: En Ano de publicação: 1995 Tipo de documento: Article