Your browser doesn't support javascript.
loading
Rapid determination of acetaminophen in plasma by LC-MS/MS / 中华劳动卫生职业病杂志
Article em Zh | WPRIM | ID: wpr-986014
Biblioteca responsável: WPRO
ABSTRACT
Objective: To establish a method for the rapid determination of acetaminophen (APAP) in human plasma by LC-MS/MS. Methods: The plasma samples were extracted by methanol and acetonitrile (1: 1) and purified directly. C(18) column was used for sample separation. The mobile phase were methanol (5 mmol/L ammonium acetate) and water (5 mmol/L ammonium acetate). Samples were analyzed by LC MS/MS with the electrospray ionization multi reaction monitoring (MRM) mode. Results: The calibration curves of APAP was linear in the concentration range of 0~10 mg/L, the correlation coefficient (r) was greater than 0.999 0. The relative standard deviation within and between batches was less than 10%. The recovery rate were 96.81%~101.7%. The detection limit of the method was 0.1 μg/L and the lower limit of quantification was 0.3 μg/L. Conclusion: This method has strong specificity, high sensitivity and reliable determination results. It is suitable for the rapid analysis of clinical plasma samples.
Assuntos
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Limite: Humans Idioma: Zh Ano de publicação: 2023 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Limite: Humans Idioma: Zh Ano de publicação: 2023 Tipo de documento: Article