RESUMEN
Model-based Roentgen Stereophotogrammetric Analysis (RSA) is able to measure the migration of metallic prostheses with submillimeter accuracy through contour-detection and 3D surface model matching techniques. However, contour-detection is only possible if the prosthesis is clearly visible in the radiograph; consequently Model-based RSA cannot be directly used for polymeric materials due to their limited X-ray attenuation; this is especially clinically relevant for all-polyethylene implants. In this study the radiopacity of unicompartmental Ultra-High Molecular Weight Polyethylene (UHMWPE) knee bearings was increased by diffusing an oil-based contrast agent into the surface to create three different levels of surface radiopacity. Model-based RSA was performed on the bearings alone, the bearings alongside a metallic component held in position using a phantom, the bearings cemented into a Sawbone tibia, and the bearings at different distances from the femoral component. For each condition the precision and accuracy of zero motion of Model-based RSA were assessed. The radiopaque bearings could be located in the stereo-radiographs using Model-based RSA an accuracy comparable to metallic parts for translational movements (0.03 mm to 0.50 mm). For rotational movements, the accuracy was lower (0.1∘ to 3.0∘). The measurement accuracy was compared for all the radiopacity levels and no significant difference was found (p=0.08). This study demonstrates that contrast enhanced radiopaque polyethylene can be used for Model-based RSA studies and has equivalent translational measurement precision to metallic parts in the superior-inferior direction.
Asunto(s)
Fotogrametría , Análisis Radioestereométrico , Fantasmas de Imagen , Polietileno/química , Polietilenos/química , Prótesis de la Rodilla , Prótesis e ImplantesRESUMEN
The aim of this study was to quantify the stability of fracture-implant complex in fractures after fixation. A total of 15 patients with an undisplaced fracture of the femoral neck, treated with either a dynamic hip screw or three cannulated hip screws, and 16 patients with an AO31-A2 trochanteric fracture treated with a dynamic hip screw or a Gamma Nail, were included. Radiostereometric analysis was used at six weeks, four months and 12 months post-operatively to evaluate shortening and rotation. Migration could be assessed in ten patients with a fracture of the femoral neck and seven with a trochanteric fracture. By four months post-operatively, a mean shortening of 5.4 mm (-0.04 to 16.1) had occurred in the fracture of the femoral neck group and 5.0 mm (-0.13 to 12.9) in the trochanteric fracture group. A wide range of rotation occurred in both types of fracture. Right-sided trochanteric fractures seem more rotationally stable than left-sided fractures. This prospective study shows that migration at the fracture site occurs continuously during the first four post-operative months, after which stabilisation occurs. This information may allow the early recognition of patients at risk of failure of fixation.