Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros

Banco de datos
Tipo del documento
Publication year range
1.
Rev Sci Instrum ; 93(10): 103703, 2022 Oct 01.
Artículo en Inglés | MEDLINE | ID: mdl-36319353

RESUMEN

A superfluid helium insert was developed for cryogenic microscopy of millimeter-sized specimens. An optical-interferometric position sensor, cryogenic objective mirror, and piezo-driven cryogenic stage were fixed to an insert holder that was immersed in superfluid helium. The single-component design stabilized the three-dimensional position of the sample, with root-mean-square deviations of (x, lateral) 0.33 nm, (y, lateral) 0.29 nm, and (z, axial) 0.25 nm. Because of the millimeter working range of the optical sensor, the working range of the sample under the active stabilization was (x, y) 5 mm and (z) 3 mm in superfluid helium at 1.8 K. The insert was used to obtain the millimeter-sized fluorescence image of cell nuclei at 1.8 K without a sample exchange.

SELECCIÓN DE REFERENCIAS
Detalles de la búsqueda