World Neurosurg ; 183: e345-e354, 2024 03.
Artículo en Inglés | MEDLINE | ID: mdl-38151174

RESUMEN

OBJECTIVE: To improve implant stability parameters, including pedicle screw (PS) outer diameter, thread depth, and pitch, by finite element analysis. METHODS: Insertion and pullout of the PS were simulated by finite element analysis, and the precision of simulation was evaluated by comparison with mechanical tests. Influences of the parameters on the maximum insertion torque and maximum pullout force were analyzed by computational simulations, including single-factor analysis and orthogonal experiments. RESULTS: The simulation results agreed with the mechanical test results. The order of parameters influencing insertion torque and pullout force was outer diameter > pitch > thread depth. When the pilot hole diameter is 0.1 mm larger than the inner diameter of the PS, the calculated Pearson correlation coefficient between the maximum insertion torque and maximum pullout force was r = 0.99. The optimized PS had a maximum insertion torque of 485.16 N·mm and a maximum pullout force of 1726.33 N, 23.9% and 9.1% higher, respectively, than the values of standard screws. CONCLUSIONS: The presently used models are feasible for evaluating the implant stability of PSs. The maximum insertion torque and maximum pullout force of PSs are highly correlated and can be improved by increasing the outer diameter and decreasing pitch. Although with the parameters of the PS, pedicle size and bone mineral density are 2 additional factors to consider for better implant stability.


Asunto(s)
Tornillos Pediculares , Humanos , Análisis de Elementos Finitos , Densidad Ósea , Torque , Fenómenos Biomecánicos , Ensayo de Materiales
SELECCIÓN DE REFERENCIAS
Detalles de la búsqueda