J Mech Behav Biomed Mater ; 94: 22-31, 2019 06.
Artigo em Inglês | MEDLINE | ID: mdl-30852348

RESUMO

We present a numerical model of a contactless test commonly used to assess the biomechanics of the human cornea. The test, consisting in a rapid air jet applied to the anterior surface of the cornea, is controversial. Although the numerous studies documented in the literature have not been able yet to clarify its relevance as a diagnostic tool, the test has the potential to be combined with inverse analysis procedures to characterize the parameters of numerical models of the cornea. With the final goal of employing the air puff test in advanced material identification algorithms, here we propose to model the cornea with standard finite elements and the fluids filling the anterior chamber of the eye with a meshfree discretization. The interaction between moving fluids and deforming cornea is accounted for by modifying the interface boundary conditions of both fluid and solid. The proposed model represents the first fully 3D example of an aqueous-cornea fluid-solid interaction analysis which uses a robust meshfree approach for the fluid. Although we restrict our scope to isotropic nonlinear materials, numerical results confirm the undeniable importance of including internal fluids in the simulation of the air puff test. Thus the proposed approach stands as a procedural paradigm for the identification of the mechanical parameters of the human cornea.


Assuntos
Ar , Córnea/fisiologia , Hidrodinâmica , Modelos Biológicos , Tonometria Ocular , Humanos , Pressão Intraocular
2.
J Mech Behav Biomed Mater ; 77: 205-216, 2018 01.
Artigo em Inglês | MEDLINE | ID: mdl-28942053

RESUMO

The air puff test is an in-vivo investigative procedure commonly utilized in ophthalmology to estimate the intraocular pressure. Potentially the test, quick and painless, could be combined with inverse analysis methods to characterize the patient-specific mechanical properties of the human cornea. A rapid localized air jet applied on the anterior surface induces the inward motion of the cornea, that interacts with aqueous humor-the fluid filling the narrow space between cornea and iris-with a strong influence on the dynamics of the cornea. While models of human cornea reproduce accurately patient-specific geometries and have reached a considerable level of complexity in the description of the material, yet scant attention has been paid to the aqueous humor, and no eye models accounting for the physically correct fluid-solid interaction are currently available. The present study addresses this gap by proposing a fluid-structure interaction approach based on a simplified two-dimensional axis-symmetric geometry to simulate the anterior chamber of the eye undergoing the air puff test. We regard the cornea as a membrane described through an analytical model and discretize the fluid with a mesh-free particle approach. The membrane is assumed to be nonlinear elastic and isotropic, and the fluid weakly compressible Newtonian. Numerical analyses reveal a marked influence of the fluid on the dynamics of the cornea. We perform a parametric analysis to assess the quantitative influence of geometrical and material parameters on the mechanical response of the model. Additionally, we investigate the possibility to use the dynamics of the test to estimate the intraocular pressure.


Assuntos
Humor Aquoso/fisiologia , Córnea/fisiologia , Pressão Intraocular , Modelos Biológicos , Tonometria Ocular/métodos , Humanos , Iris/fisiologia , Movimento (Física) , Pressão , Viscosidade
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa