Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 7 de 7
Filtrar
1.
Phys Med Biol ; 54(17): N375-83, 2009 Sep 07.
Artigo em Inglês | MEDLINE | ID: mdl-19652292

RESUMO

This study presents a method aimed at creating radiotherapy (RT) patient-adjustable whole-body phantoms to permit retrospective and prospective peripheral dose evaluations for enhanced patient radioprotection. Our strategy involves virtual whole-body patient models (WBPM) in different RT treatment positions for both genders and for different age groups. It includes a software tool designed to match the anatomy of the phantoms with the anatomy of the actual patients, based on the quality of patient data available. The procedure for adjusting a WBPM to patient morphology includes typical dimensions available in basic auxological tables for the French population. Adjustment is semi-automatic. Because of the complexity of the human anatomy, skilled personnel are required to validate changes made in the phantom anatomy. This research is part of a global project aimed at proposing appropriate methods and software tools capable of reconstituting the anatomy and dose evaluations in the entire body of RT patients in an adapted treatment planning system (TPS). The graphic user interface is that of a TPS adapted to obtain a comfortable working process. Such WBPM have been used to supplement patient therapy planning images, usually restricted to regions involved in treatment. Here we report, as an example, the case of a patient treated for prostate cancer whose therapy planning images were complemented by an anatomy model. Although present results are preliminary and our research is ongoing, they appear encouraging, since such patient-adjusted phantoms are crucial in the optimization of radiation protection of patients and for follow-up studies.


Assuntos
Imagens de Fantasmas , Radiometria/instrumentação , Radioterapia/métodos , Software , Adolescente , Adulto , Criança , Pré-Escolar , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Postura , Radiometria/métodos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
2.
Cancer Radiother ; 12(8): 800-8, 2008 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-18571453

RESUMO

PURPOSE: The aim of the present study was to quantitatively assess the performance of a block matching-based automatic registration algorithm integrated within the commercial treatment planning system designated ISOgray from Dosisoft. The accuracy of the process was evaluated by a phantom study on computed tomography (CT), magnetic resonance (MR) and positron emission tomography (PET) images. MATERIALS AND METHODS: Two phantoms were used to carry out this study: the cylindrical Jaszczak phantom and the anthropomorphic Liqui-Phil Head Phantom (the Phantom Laboratory), containing fillable spheres. External fiducial markers were used to quantify the accuracy of 41 CT/CT, MR/CT and PET/CT automatic registrations with images of the rotated and tilted phantoms. RESULTS: The study first showed that a cylindrical phantom was not adapted for the evaluation of the performance of a block matching-based registration software. Secondly, the Liqui-Phil Head Phantom study showed that the algorithm was able to perform automatic registrations of CT/CT and MR/CT images with differences of up to 40 degrees in phantom rotation and of up to 20-30 degrees for PET/CT with accuracy below the image voxel size. CONCLUSION: The study showed that the block matching-based automatic registration software under investigation was robust, reliable and yielded very satisfactory results. This phantom-based test can be integrated into a periodical quality assurance process and used for any commissioning of image registration software for radiation therapy.


Assuntos
Imageamento por Ressonância Magnética/métodos , Imagens de Fantasmas , Tomografia por Emissão de Pósitrons/métodos , Tomografia Computadorizada por Raios X/métodos , Algoritmos , Automação , Encéfalo/diagnóstico por imagem , Encéfalo/patologia , Cabeça , Humanos , Planejamento da Radioterapia Assistida por Computador/métodos , Planejamento da Radioterapia Assistida por Computador/normas , Software
3.
Cancer Radiother ; 12(8): 781-7, 2008 Dec.
Artigo em Francês | MEDLINE | ID: mdl-18691926

RESUMO

PURPOSE: This paper presents the development of a protocol for quality control of a megavoltage cone beam CT imaging system (MVCB) mounted on a Siemens Oncor 6MV linear accelerator. MATERIALS AND METHODS: Several parameters were controlled on the MVCB system: (1) the initial geometric calibration of the system; (2) the quality of the images (geometric distortion, uniformity, spatial resolution, low contrast resolution) for various protocols; (3) the correspondence between the intensity of voxels and electronic densities; (4) the dose delivered when achieving a MVCB. These tests were done mainly with two cylindrical phantoms specific to the quality control (QC) of a MVCB system, supplied by Siemens, and with the Catphan 600 phantom (The Phantom Laboratory) and Quasar Multipurpose Body phantom (Modus Medical Devices Inc). RESULTS: The results of the quality control of the images were within the tolerances. The use of the Catphan 600 phantom was inadequate for the QC of MVCB images. These tests also highlighted the need to correct the MVCB images for the "cupping artefact" for dose calculation purpose. CONCLUSION: The initial characteristics of the MVCB imaging system were established. Such testing also provided the assessment of the influence of various parameters on the image quality as well as the associated dose delivered during their acquisition, and emphasized the corrections needed to use MVCB images for dose calculation.


Assuntos
Aceleradores de Partículas/normas , Tomografia Computadorizada por Raios X/normas , Calibragem , Humanos , Imagens de Fantasmas , Controle de Qualidade , Doses de Radiação , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
4.
Phys Med ; 38: 16-22, 2017 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-28610693

RESUMO

OBJECTIVE: To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure. METHODS: The study was conducted on a total of 40 cardiac procedures (20 complex coronary angioplasty of chronic total occlusion (CTO) and 20 coronary angiography and coronary angioplasty (CA-PTCA)) conducted between January 2014 to December 2015. PSD measurement (PSDFilm) was obtained by placing XR-RV3 Gafchromic under the patient's back for each procedure. PSD (PSDem.dose) was computed with the software em.dose©. The calculation was performed on the dose metrics collected from the private dose report of each procedure. Two calculation methods (method A: fluoroscopic kerma equally spread on cine acquisition and B: fluoroscopic kerma is added to one air Kerma cine acquisition that contributes to the PSD) were used to calculate the fluoroscopic dose contribution as fluoroscopic data were not recorded in our interventional room. Statistical analyses were carried out to compare PSDFilm and PSDem.dose. RESULTS: The PSDFilm median (1st quartile; 3rd quartile) was 0.251(0.190;0.336)Gy for CA-PTCA and 1.453(0.767;2.011)Gy for CTO. For method-A, the PSDem.dose was 0.248(0.182;0.369)Gy for CA-PTCA and 1.601(0.892;2.178)Gy for CTO, and 0.267(0.223;0.446)Gy and 1.75 (0.912;2.584)Gy for method-B, respectively. For the two methods, the correlation between PSDFilm and PSDem.dose was strong. For all cardiology procedures investigated, the mean deviation between PSDFilm and PSDem.dose was 3.4±21.1% for method-A and 17.3%±23.9% for method-B. CONCLUSION: The dose mapping software is convenient to calculate peak skin dose in interventional cardiology.


Assuntos
Doses de Radiação , Pele/efeitos da radiação , Software , Adulto , Idoso , Idoso de 80 Anos ou mais , Angioplastia , Cardiologia/métodos , Angiografia Coronária , Feminino , Fluoroscopia , Humanos , Masculino , Pessoa de Meia-Idade , Dosímetros de Radiação , Radiometria
5.
Radiat Prot Dosimetry ; 174(2): 207-215, 2017 Apr 25.
Artigo em Inglês | MEDLINE | ID: mdl-27150522

RESUMO

In this study, the impact of radiochromic films' (XR-RV3) calibration on PSD measurements was investigated under various peak kilovoltage (kVp) and additional filtration conditions. Films were calibrated free-in-air for six beam qualities with Allura Xper FD20 system (Philips). Six calibration curves (CCs) were constructed. Each beam quality was characterized in terms of mean energy (ME) in the air, with table, with table and water phantom using Monte Carlo simulations. A cohort of 155 patient films from cardiology (37) and vascular (118) procedures were read with each CC. Routine calibration beam quality was taken as reference (DoseNorm). Overall, it was observed that for a wider ME difference between the exposed film and the CC used, a larger deviation (from -28% to +41%) was observed. The choice of beam quality for the calibration is a key point when additional filtration and kVp are automatically controlled in clinical conditions.


Assuntos
Método de Monte Carlo , Radiologia Intervencionista , Calibragem , Dosimetria Fotográfica , Humanos , Imagens de Fantasmas
6.
J Clin Endocrinol Metab ; 91(3): 920-5, 2006 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-16368753

RESUMO

OBJECTIVE: Patients with adrenocortical cancer are submitted to multiple imaging procedures for diagnosis of recurrence and staging. The aim of this prospective study was to evaluate the diagnostic and prognostic values of fluorodeoxyglucose (FDG) using a combined positron emission tomography and computed tomography (PET/CT) modality, compared with thoracoabdominopelvic computed tomography (TAP-CT). METHODS: Twenty-eight consecutive patients with adrenocortical cancer referred from November 2003 to December 2004 to the Institut Gustave Roussy were included. Mean time between PET/CT and TAP-CT was 16 d. Independent readers analyzed images of each modality. The gold standard was progression on follow-up TAP-CT or pathology. RESULTS: A total of 269 lesions in 57 organs were depicted in 22 patients. The sensitivities for the detection of distinct lesions and the diagnosis of metastatic organs were 90 and 93% for PET/CT and 88 and 82% for TAP-CT, respectively. Twelve percent of the lesions were seen on PET/CT only and 10% on TAP-CT only. Eighteen percent of the metastatic organs were diagnosed with PET/CT only and 7% with TAP-CT only. Thirty-eight percent of the local relapses were seen only with PET/CT. PET/CT depicted three false-positive lesions. Treatment modalities were modified by PET/CT findings in five cases among which one was falsely positive. Tumor size and mitotic rate were significantly associated with FDG uptake. The intensity of FDG uptake (maximum standardized uptake value > 10) and the volume of FDG uptake (>150 ml) were significant prognostic factors for survival. CONCLUSIONS: We show that FDG-PET/CT is complementary to TAP-CT and of special interest in the diagnosis of local relapses.


Assuntos
Neoplasias do Córtex Suprarrenal/diagnóstico por imagem , Fluordesoxiglucose F18 , Neoplasias do Córtex Suprarrenal/mortalidade , Neoplasias do Córtex Suprarrenal/patologia , Adulto , Idoso , Feminino , Humanos , Processamento de Imagem Assistida por Computador , Masculino , Pessoa de Meia-Idade , Metástase Neoplásica , Tomografia por Emissão de Pósitrons , Prognóstico , Compostos Radiofarmacêuticos , Análise de Sobrevida , Tomografia Computadorizada por Raios X
7.
Cancer Radiother ; 10(5): 222-30, 2006 Sep.
Artigo em Francês | MEDLINE | ID: mdl-16890471

RESUMO

The development of conformal radiotherapy techniques (CRT) and intensity modulated CRT requires an accurate delineation of target structures and organs at risk. Thus, additional information provided by anatomical and/or functional imaging modalities can be used for volume of interest determination combined with traditionally used Computed Tomography imaging (CT): for instance, functional or morphological Magnetic Resonance Imaging (f MRI or m MRI) or Positron Emission Tomography (PET). A prerequisite to the simultaneous use of this information is image registration. Due to the differences between the images and the information they provide, a quality control of image registration process for radiotherapy is mandatory. The purpose of this article is to present the difficulties in implementing such controls and to show the necessity for a clinical validation on patient's images. The last part of this work presents the possible interest in using f MRI to help radio-oncologists in the treatment planning for gliomas associated to image coregistration and quality control considerations.


Assuntos
Diagnóstico por Imagem/métodos , Processamento de Imagem Assistida por Computador/métodos , Planejamento da Radioterapia Assistida por Computador , Radioterapia Assistida por Computador , Neoplasias Encefálicas/radioterapia , Diagnóstico por Imagem/normas , Glioma/radioterapia , Humanos , Processamento de Imagem Assistida por Computador/normas , Imageamento por Ressonância Magnética , Tomografia por Emissão de Pósitrons , Controle de Qualidade , Planejamento da Radioterapia Assistida por Computador/métodos , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia Assistida por Computador/métodos , Radioterapia Assistida por Computador/normas , Radioterapia Conformacional , Radioterapia de Intensidade Modulada , Tomografia Computadorizada por Raios X
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa