Lab Med ; 54(2): e44-e48, 2023 Mar 07.
Artigo em Inglês | MEDLINE | ID: mdl-35976956

RESUMO

OBJECTIVE: The purpose of this study was to compare Quidel's rapid antigen test Sofia SARS antigen Fluorescent Immunoassay (FIA) (Sofia) with the real-time reverse transcription-polymerase chain reaction (rRT-PCR) test. METHODS: Two samples were taken from each test subject-1 for testing with the Sofia test and 1 for testing with the rRT-PCR test. In total, swabs were taken from 146 subjects who presented symptoms of infection (group 1) and 672 subjects who were tested regardless of symptoms (group 2). RESULTS: In group 1, the sensitivity of the antigen test was 90.0% and its specificity 97.5%. In group 2, however, the sensitivity of the antigen test was 81.4% and the specificity 98.9%. In addition to asymptomatic patients, false-negative results of rapid antigen tests also occurred in subjects with high threshold values (cycle threshold > 30). CONCLUSION: Our results show that the Sofia test meets the standards for diagnostic tests according to the criteria of the World Health Organization, as they show high sensitivity and specificity, and perhaps most importantly, a high negative predictive value (> 95%).


Assuntos
COVID-19 , Humanos , COVID-19/diagnóstico , Sensibilidade e Especificidade , SARS-CoV-2/genética , Valor Preditivo dos Testes , Reação em Cadeia da Polimerase em Tempo Real , Teste para COVID-19