BACKGROUND: Surges of COVID-19 infections have led to insufficient supply of mechanical ventilators (MV), resulting in rationing of MV care. In-parallel, co-mechanical ventilation (Co-MV) of multiple patients is a potential solution. However, due to lack of testing, there is currently no means to match ventilation requirements or patients, with no guidelines to date. In this research, we have developed a model-based method for patient matching for pressure control mode MV. METHODS: The model-based method uses a single-compartment lung model (SCM) to simulate the resultant tidal volume of patient pairs at a set ventilation setting. If both patients meet specified safe ventilation criteria under similar ventilation settings, the actual mechanical ventilator settings for Co-MV are determined via simulation using a double-compartment lung model (DCM). This method allows clinicians to analyse Co-MV in silico, before clinical implementation. RESULTS: The proposed method demonstrates successful patient matching and MV setting in a model-based simulation as well as good discrimination to avoid mismatched patient pairs. The pairing process is based on model-based, patient-specific respiratory mechanics identified from measured data to provide useful information for guiding care. Specifically, the matching is performed via estimation of MV delivered tidal volume (mL/kg) based on patient-specific respiratory mechanics. This information can provide insights for the clinicians to evaluate the subsequent effects of Co-MV. In addition, it was also found that Co-MV patients with highly restrictive respiratory mechanics and obese patients must be performed with extra care. CONCLUSION: This approach allows clinicians to analyse patient matching in a virtual environment without patient risk. The approach is tested in simulation, but the results justify the necessary clinical validation in human trials.
Assuntos
COVID-19 , Humanos , Respiração Artificial , SARS-CoV-2 , Volume de Ventilação Pulmonar , Ventiladores MecânicosRESUMO
BACKGROUND AND OBJECTIVE: Mechanical ventilation (MV) is the primary form of care for respiratory failure patients. MV settings are based on general clinical guidelines, intuition, and experience. This approach is not patient-specific and patients may thus experience suboptimal, potentially harmful MV care. This study presents the Stochastic integrated VENT (SiVENT) protocol which combines model-based approaches of the VENT protocol from previous works, with stochastic modelling to take the variation of patient respiratory elastance over time into consideration. METHODS: A stochastic model of Ers is integrated into the VENT protocol from previous works to develop the SiVENT protocol, to account for both intra- and inter-patient variability. A cohort of 20 virtual MV patients based on retrospective patient data are used to validate the performance of this method for volume-controlled (VC) ventilation. A performance evaluation was conducted where the SiVENT and VENT protocols were implemented in 1080 instances each to compare the two protocols and evaluate the difference in reduction of possible MV settings achieved by each. RESULTS: From an initial number of 189,000 possible MV setting combinations, the VENT protocol reduced this number to a median of 10,612, achieving a reduction of 94.4% across the cohort. With the integration of the stochastic model component, the SiVENT protocol reduced this number from 189,000 to a median of 9329, achieving a reduction of 95.1% across the cohort. The SiVENT protocol reduces the number of possible combinations provided to the user by more than 1000 combinations as compared to the VENT protocol. CONCLUSIONS: Adding a stochastic model component into a model-based approach to selecting MV settings improves the ability of a decision support system to recommend patient-specific MV settings. It specifically considers inter- and intra-patient variability in respiratory elastance and eliminates potentially harmful settings based on clinically recommended pressure thresholds. Clinical input and local protocols can further reduce the number of safe setting combinations. The results for the SiVENT protocol justify further investigation of its prediction accuracy and clinical validation trials.
Assuntos
Respiração Artificial , Sistema Respiratório , Humanos , Estudos RetrospectivosRESUMO
BACKGROUND: Patient-specific lung mechanics during mechanical ventilation (MV) can be identified from measured waveforms of fully ventilated, sedated patients. However, asynchrony due to spontaneous breathing (SB) effort can be common, altering these waveforms and reducing the accuracy of identified, model-based, and patient-specific lung mechanics. METHODS: Changes in patient-specific lung elastance over a pressure-volume (PV) loop, identified using hysteresis loop analysis (HLA), are used to detect the occurrence of asynchrony and identify its type and pattern. The identified HLA parameters are then combined with a nonlinear mechanics hysteresis loop model (HLM) to extract and reconstruct ventilated waveforms unaffected by asynchronous breaths. Asynchrony magnitude can then be quantified using an energy-dissipation metric, Easyn, comparing PV loop area between model-reconstructed and original, altered asynchronous breathing cycles. Performance is evaluated using both test-lung experimental data with a known ground truth and clinical data from four patients with varying levels of asynchrony. RESULTS: Root mean square errors for reconstructed PV loops are within 5% for test-lung experimental data, and 10% for over 90% of clinical data. Easyn clearly matches known asynchrony magnitude for experimental data with RMS errors < 4.1%. Clinical data performance shows 57% breaths having Easyn > 50% for Patient 1 and 13% for Patient 2. Patient 3 only presents 20% breaths with Easyn > 10%. Patient 4 has Easyn = 0 for 96% breaths showing accuracy in a case without asynchrony. CONCLUSIONS: Experimental test-lung validation demonstrates the method's reconstruction accuracy and generality in controlled scenarios. Clinical validation matches direct observations of asynchrony in incidence and quantifies magnitude, including cases without asynchrony, validating its robustness and potential efficacy as a clinical real-time asynchrony monitoring tool.
Assuntos
Respiração Artificial , Mecânica Respiratória , Humanos , Modelos Biológicos , Dinâmica não Linear , Testes de Função Respiratória , Mecânica Respiratória/fisiologiaRESUMO
BACKGROUND: Mechanical ventilation is an essential therapy to support critically ill respiratory failure patients. Current standards of care consist of generalised approaches, such as the use of positive end expiratory pressure to inspired oxygen fraction (PEEP-FiO2) tables, which fail to account for the inter- and intra-patient variability between and within patients. The benefits of higher or lower tidal volume, PEEP, and other settings are highly debated and no consensus has been reached. Moreover, clinicians implicitly account for patient-specific factors such as disease condition and progression as they manually titrate ventilator settings. Hence, care is highly variable and potentially often non-optimal. These conditions create a situation that could benefit greatly from an engineered approach. The overall goal is a review of ventilation that is accessible to both clinicians and engineers, to bridge the divide between the two fields and enable collaboration to improve patient care and outcomes. This review does not take the form of a typical systematic review. Instead, it defines the standard terminology and introduces key clinical and biomedical measurements before introducing the key clinical studies and their influence in clinical practice which in turn flows into the needs and requirements around how biomedical engineering research can play a role in improving care. Given the significant clinical research to date and its impact on this complex area of care, this review thus provides a tutorial introduction around the review of the state of the art relevant to a biomedical engineering perspective. DISCUSSION: This review presents the significant clinical aspects and variables of ventilation management, the potential risks associated with suboptimal ventilation management, and a review of the major recent attempts to improve ventilation in the context of these variables. The unique aspect of this review is a focus on these key elements relevant to engineering new approaches. In particular, the need for ventilation strategies which consider, and directly account for, the significant differences in patient condition, disease etiology, and progression within patients is demonstrated with the subsequent requirement for optimal ventilation strategies to titrate for patient- and time-specific conditions. CONCLUSION: Engineered, protective lung strategies that can directly account for and manage inter- and intra-patient variability thus offer great potential to improve both individual care, as well as cohort clinical outcomes.
Assuntos
Engenharia Biomédica , Cuidados Críticos , Respiração com Pressão Positiva/instrumentação , Respiração Artificial/instrumentação , Animais , Estado Terminal , Humanos , Pulmão , Lesão Pulmonar/etiologia , Oscilometria , Oxigênio/sangue , Oxigênio/química , Respiração com Pressão Positiva/métodos , Pressão , Respiração Artificial/métodos , Síndrome do Desconforto Respiratório/terapia , Risco , Volume de Ventilação Pulmonar , Ventiladores MecânicosRESUMO
Critical care, like many healthcare areas, is under a dual assault from significantly increasing demographic and economic pressures. Intensive care unit (ICU) patients are highly variable in response to treatment, and increasingly aging populations mean ICUs are under increasing demand and their cohorts are increasingly ill. Equally, patient expectations are growing, while the economic ability to deliver care to all is declining. Better, more productive care is thus the big challenge. One means to that end is personalised care designed to manage the significant inter- and intra-patient variability that makes the ICU patient difficult. Thus, moving from current "one size fits all" protocolised care to adaptive, model-based "one method fits all" personalised care could deliver the required step change in the quality, and simultaneously the productivity and cost, of care. Computer models of human physiology are a unique tool to personalise care, as they can couple clinical data with mathematical methods to create subject-specific models and virtual patients to design new, personalised and more optimal protocols, as well as to guide care in real-time. They rely on identifying time varying patient-specific parameters in the model that capture inter- and intra-patient variability, the difference between patients and the evolution of patient condition. Properly validated, virtual patients represent the real patients, and can be used in silico to test different protocols or interventions, or in real-time to guide care. Hence, the underlying models and methods create the foundation for next generation care, as well as a tool for safely and rapidly developing personalised treatment protocols over large virtual cohorts using virtual trials. This review examines the models and methods used to create virtual patients. Specifically, it presents the models types and structures used and the data required. It then covers how to validate the resulting virtual patients and trials, and how these virtual trials can help design and optimise clinical trial. Links between these models and higher order, more complex physiome models are also discussed. In each section, it explores the progress reported up to date, especially on core ICU therapies in glycemic, circulatory and mechanical ventilation management, where high cost and frequency of occurrence provide a significant opportunity for model-based methods to have measurable clinical and economic impact. The outcomes are readily generalised to other areas of medical care.
Assuntos
Simulação por Computador , Cuidados Críticos/métodos , Modelos Biológicos , Medicina de Precisão/métodos , Estudos de Coortes , Humanos , Fenômenos FisiológicosAssuntos
Betacoronavirus , Infecções por Coronavirus , Pandemias , Pneumonia Viral , COVID-19 , Humanos , SARS-CoV-2 , Ventiladores MecânicosAssuntos
Betacoronavirus , Infecções por Coronavirus , Pandemias , Pneumonia Viral , COVID-19 , Humanos , SARS-CoV-2RESUMO
BACKGROUND: Real-time patient respiratory mechanics estimation can be used to guide mechanical ventilation settings, particularly, positive end-expiratory pressure (PEEP). This work presents a software, Clinical Utilisation of Respiratory Elastance (CURE Soft), using a time-varying respiratory elastance model to offer this ability to aid in mechanical ventilation treatment. IMPLEMENTATION: CURE Soft is a desktop application developed in JAVA. It has two modes of operation, 1) Online real-time monitoring decision support and, 2) Offline for user education purposes, auditing, or reviewing patient care. The CURE Soft has been tested in mechanically ventilated patients with respiratory failure. The clinical protocol, software testing and use of the data were approved by the New Zealand Southern Regional Ethics Committee. RESULTS AND DISCUSSION: Using CURE Soft, patient's respiratory mechanics response to treatment and clinical protocol were monitored. Results showed that the patient's respiratory elastance (Stiffness) changed with the use of muscle relaxants, and responded differently to ventilator settings. This information can be used to guide mechanical ventilation therapy and titrate optimal ventilator PEEP. CONCLUSION: CURE Soft enables real-time calculation of model-based respiratory mechanics for mechanically ventilated patients. Results showed that the system is able to provide detailed, previously unavailable information on patient-specific respiratory mechanics and response to therapy in real-time. The additional insight available to clinicians provides the potential for improved decision-making, and thus improved patient care and outcomes.
Assuntos
Mecânica Respiratória/fisiologia , Software , Humanos , Respiração com Pressão Positiva/métodos , Respiração Artificial/métodos , Ventiladores MecânicosRESUMO
BACKGROUND: Patients with acute respiratory distress syndrome (ARDS) risk lung collapse, severely altering the breath-to-breath respiratory mechanics. Model-based estimation of respiratory mechanics characterising patient-specific condition and response to treatment may be used to guide mechanical ventilation (MV). This study presents a model-based approach to monitor time-varying patient-ventilator interaction to guide positive end expiratory pressure (PEEP) selection. METHODS: The single compartment lung model was extended to monitor dynamic time-varying respiratory system elastance, Edrs, within each breathing cycle. Two separate animal models were considered, each consisting of three fully sedated pure pietrain piglets (oleic acid ARDS and lavage ARDS). A staircase recruitment manoeuvre was performed on all six subjects after ARDS was induced. The Edrs was mapped across each breathing cycle for each subject. RESULTS: Six time-varying, breath-specific Edrs maps were generated, one for each subject. Each Edrs map shows the subject-specific response to mechanical ventilation (MV), indicating the need for a model-based approach to guide MV. This method of visualisation provides high resolution insight into the time-varying respiratory mechanics to aid clinical decision making. Using the Edrs maps, minimal time-varying elastance was identified, which can be used to select optimal PEEP. CONCLUSIONS: Real-time continuous monitoring of in-breath mechanics provides further insight into lung physiology. Therefore, there is potential for this new monitoring method to aid clinicians in guiding MV treatment. These are the first such maps generated and they thus show unique results in high resolution. The model is limited to a constant respiratory resistance throughout inspiration which may not be valid in some cases. However, trends match clinical expectation and the results highlight both the subject-specificity of the model, as well as significant inter-subject variability.
Assuntos
Síndrome do Desconforto Respiratório/fisiopatologia , Mecânica Respiratória , Animais , Modelos Animais de Doenças , Suínos , Fatores de TempoRESUMO
Sonothrombolysis is a technique that employs the ultrasound waves to break down the clot. Recent studies have demonstrated significant improvement in the treatment efficacy when combining two ultrasound waves of different frequencies. Nevertheless, the findings remain conflicted on the ideal frequency pairing that leads to an optimal treatment outcome. Existing experimental studies are constrained by the limited range of frequencies that can be investigated, while numerical studies are typically confined to spherical microbubble dynamics, thereby restricting the scope of the analysis. To overcome this, the present study investigated the microbubble dynamics caused by the different combinations of ultrasound frequencies. This was carried out using computational modelling as it enables the visualisation of the microbubble behaviour, which is difficult in experimental studies due to the opacity of blood. The results showed that the pairings of two ultrasound waves with low frequencies generally produced stronger cavitation and higher flow-induced shear stress on the clot surface. However, one should avoid the frequency pairings that are integer multipliers of each other, i.e., frequency ratio of 1/3, 1/2 and 2, as they led to resultant wave with low pressure amplitude that weakened the cavitation. At 0.5 + 0.85 MHz, the microbubble caused the highest shear stress of 60.5 kPa, due to its large translational distance towards the clot. Although the pressure threshold for inertial cavitation was reduced using dual-frequency ultrasound, the impact of the high-speed jet can only be realised when the microbubble travelled close to the clot. The results obtained from the present study provide groundwork for deeper understanding on the microbubble dynamics during dual-frequency sonothrombolysis, which is of paramount importance for its optimisations and the subsequent clinical translation.
Assuntos
Simulação por Computador , Microbolhas , Terapia por Ultrassom , Humanos , Terapia por Ultrassom/métodos , Modelos Cardiovasculares , Trombose/diagnóstico por imagemRESUMO
BACKGROUND AND OBJECTIVE: Patient-ventilator asynchrony (PVA) is associated with poor clinical outcomes and remains under-monitored. Automated PVA detection would enable complete monitoring standard observational methods do not allow. While model-based and machine learning PVA approaches exist, they have variable performance and can miss specific PVA events. This study compares a model and rule-based algorithm with a machine learning PVA method by retrospectively validating both methods using an independent patient cohort. METHODS: Hysteresis loop analysis (HLA) which is a rule-based method (RBM) and a tri-input convolutional neural network (TCNN) machine learning model are used to classify 7 different types of PVA, including: 1) flow asynchrony; 2) reverse triggering; 3) premature cycling; 4) double triggering; 5) delayed cycling; 6) ineffective efforts; and 7) auto triggering. Class activation mapping (CAM) heatmaps visualise sections of respiratory waveforms the TCNN model uses for decision making, improving result interpretability. Both PVA classification methods were used to classify incidence in an independent retrospective clinical cohort of 11 mechanically ventilated patients for validation and performance comparison. RESULTS: Self-validation with the training dataset shows overall better HLA performance (accuracy, sensitivity, specificity: 97.5 %, 96.6 %, 98.1 %) compared to the TCNN model (accuracy, sensitivity, specificity: 89.5 %, 98.3 %, 83.9 %). In this study, the TCNN model demonstrates higher sensitivity in detecting PVA, but HLA was better at identifying non-PVA breathing cycles due to its rule-based nature. While the overall AI identified by both classification methods are very similar, the intra-patient distribution of each PVA type varies between HLA and TCNN. CONCLUSION: The collective findings underscore the efficacy of both HLA and TCNN in PVA detection, indicating the potential for real-time continuous monitoring of PVA. While ML methods such as TCNN demonstrate good PVA identification performance, it is essential to ensure optimal model architecture and diversity in training data before widespread uptake as standard care. Moving forward, further validation and adoption of RBM methods, such as HLA, offers an effective approach to PVA detection while providing clear distinction into the underlying patterns of PVA, better aligning with clinical needs for transparency, explicability, adaptability and reliability of these emerging tools for clinical care.
Assuntos
Algoritmos , Aprendizado de Máquina , Redes Neurais de Computação , Respiração Artificial , Humanos , Estudos Retrospectivos , Masculino , Feminino , Pessoa de Meia-Idade , Idoso , Ventiladores Mecânicos , Assincronia Paciente-VentiladorRESUMO
BACKGROUND: Acute Respiratory Distress Syndrome (ARDS) is characterized by inflammation, filling of the lung with fluid and the collapse of lung units. Mechanical ventilation (MV) is used to treat ARDS using positive end expiratory pressure (PEEP) to recruit and retain lung units, thus increasing pulmonary volume and dynamic functional residual capacity (dFRC) at the end of expiration. However, simple, non-invasive methods to estimate dFRC do not exist. METHODS: Four model-based methods for estimating dFRC are compared based on their performance on two separate clinical data cohorts. The methods are derived from either stress-strain theory or a single compartment lung model, and use commonly controlled or measured parameters (lung compliance, plateau airway pressure, pressure-volume (PV) data). Population constants are determined for the stress-strain approach, which is implemented using data at both single and multiple PEEP levels. Estimated values are compared to clinically measured values to assess the reliability of each method for each cohort individually and combined. RESULTS: The stress-strain multiple breath (at multiple PEEP levels) method produced an overall correlation coefficient R2 = 0.966. The stress-strain single breath method produced R2 = 0.530. The single compartment single breath method produced R2 = 0.415. A combined method at single and multiple PEEP levels produced R2 = 0.963. CONCLUSIONS: The results suggest that model-based, single breath and non-invasive approaches to estimating dFRC may be viable in a clinical scenario, ensuring no interruption to MV. The models provide a means of estimating dFRC at any PEEP level. However, model limitations and large estimation errors limit the use of the methods at very low PEEP.
Assuntos
Capacidade Residual Funcional , Modelos Biológicos , Adulto , Idoso , Idoso de 80 Anos ou mais , Feminino , Humanos , Pulmão/fisiopatologia , Medidas de Volume Pulmonar , Masculino , Pessoa de Meia-Idade , Respiração com Pressão Positiva/métodos , Reprodutibilidade dos Testes , Respiração Artificial/métodos , Síndrome do Desconforto Respiratório/fisiopatologia , Síndrome do Desconforto Respiratório/terapia , Estudos Retrospectivos , Adulto JovemRESUMO
INTRODUCTION: Model-based methods can be used to characterise patient-specific condition and response to mechanical ventilation (MV) during treatment for acute respiratory distress syndrome (ARDS). Conventional metrics of respiratory mechanics are based on inspiration only, neglecting data from the expiration cycle. However, it is hypothesised that expiratory data can be used to determine an alternative metric, offering another means to track patient condition and guide positive end expiratory pressure (PEEP) selection. METHODS: Three fully sedated, oleic acid induced ARDS piglets underwent three experimental phases. Phase 1 was a healthy state recruitment manoeuvre. Phase 2 was a progression from a healthy state to an oleic acid induced ARDS state. Phase 3 was an ARDS state recruitment manoeuvre. The expiratory time-constant model parameter was determined for every breathing cycle for each subject. Trends were compared to estimates of lung elastance determined by means of an end-inspiratory pause method and an integral-based method. All experimental procedures, protocols and the use of data in this study were reviewed and approved by the Ethics Committee of the University of Liege Medical Faculty. RESULTS: The overall median absolute percentage fitting error for the expiratory time-constant model across all three phases was less than 10 %; for each subject, indicating the capability of the model to capture the mechanics of breathing during expiration. Provided the respiratory resistance was constant, the model was able to adequately identify trends and fundamental changes in respiratory mechanics. CONCLUSION: Overall, this is a proof of concept study that shows the potential of continuous monitoring of respiratory mechanics in clinical practice. Respiratory system mechanics vary with disease state development and in response to MV settings. Therefore, titrating PEEP to minimal elastance theoretically results in optimal PEEP selection. Trends matched clinical expectation demonstrating robustness and potential for guiding MV therapy. However, further research is required to confirm the use of such real-time methods in actual ARDS patients, both sedated and spontaneously breathing.
Assuntos
Expiração , Modelos Biológicos , Síndrome do Desconforto Respiratório/fisiopatologia , Progressão da Doença , Humanos , Medicina de Precisão , Respiração Artificial , Síndrome do Desconforto Respiratório/terapia , Fatores de TempoRESUMO
BACKGROUND: Neurally adjusted ventilatory assist (NAVA) delivers pressure in proportion to diaphragm electrical activity (Eadi). However, each patient responds differently to NAVA levels. This study aims to examine the matching between tidal volume (Vt) and patients' inspiratory demand (Eadi), and to investigate patient-specific response to various NAVA levels in non-invasively ventilated patients. METHODS: 12 patients were ventilated non-invasively with NAVA using three different NAVA levels. NAVA100 was set according to the manufacturer's recommendation to have similar peak airway pressure as during pressure support. NAVA level was then adjusted ±50% (NAVA50, NAVA150). Airway pressure, flow and Eadi were recorded for 15 minutes at each NAVA level. The matching of Vt and integral of Eadi (ÊEadi) were assessed at the different NAVA levels. A metric, Range90, was defined as the 5-95% range of Vt/ÊEadi ratio to assess matching for each NAVA level. Smaller Range90 values indicated better matching of supply to demand. RESULTS: Patients ventilated at NAVA50 had the lowest Range90 with median 25.6 uVs/ml [Interquartile range (IQR): 15.4-70.4], suggesting that, globally, NAVA50 provided better matching between ÊEadi and Vt than NAVA100 and NAVA150. However, on a per-patient basis, 4 patients had the lowest Range90 values in NAVA100, 1 patient at NAVA150 and 7 patients at NAVA50. Robust coefficient of variation for ÊEadi and Vt were not different between NAVA levels. CONCLUSIONS: The patient-specific matching between ÊEadi and Vt was variable, indicating that to obtain the best possible matching, NAVA level setting should be patient specific. The Range90 concept presented to evaluate Vt/ÊEadi is a physiologic metric that could help in individual titration of NAVA level.
Assuntos
Diafragma/fisiopatologia , Fenômenos Eletrofisiológicos , Suporte Ventilatório Interativo/métodos , Idoso , Feminino , Humanos , Inalação/fisiologia , Masculino , Pessoa de Meia-Idade , Medicina de Precisão , Volume de Ventilação PulmonarRESUMO
Neurally adjusted ventilatory assist (NAVA) is a ventilation assist mode that delivers pressure in proportionality to electrical activity of the diaphragm (Eadi). Compared to pressure support ventilation (PS), it improves patient-ventilator synchrony and should allow a better expression of patient's intrinsic respiratory variability. We hypothesize that NAVA provides better matching in ventilator tidal volume (Vt) to patients inspiratory demand. 22 patients with acute respiratory failure, ventilated with PS were included in the study. A comparative study was carried out between PS and NAVA, with NAVA gain ensuring the same peak airway pressure as PS. Robust coefficients of variation (CVR) for Eadi and Vt were compared for each mode. The integral of Eadi (ÊEadi) was used to represent patient's inspiratory demand. To evaluate tidal volume and patient's demand matching, Range90 = 5-95 % range of the Vt/ÊEadi ratio was calculated, to normalize and compare differences in demand within and between patients and modes. In this study, peak Eadi and ÊEadi are correlated with median correlation of coefficients, R > 0.95. Median ÊEadi, Vt, neural inspiratory time (Ti_ ( Neural )), inspiratory time (Ti) and peak inspiratory pressure (PIP) were similar in PS and NAVA. However, it was found that individual patients have higher or smaller ÊEadi, Vt, Ti_ ( Neural ), Ti and PIP. CVR analysis showed greater Vt variability for NAVA (p < 0.005). Range90 was lower for NAVA than PS for 21 of 22 patients. NAVA provided better matching of Vt to ÊEadi for 21 of 22 patients, and provided greater variability Vt. These results were achieved regardless of differences in ventilatory demand (Eadi) between patients and modes.
Assuntos
Diafragma/fisiologia , Eletromiografia , Suporte Ventilatório Interativo/métodos , Respiração com Pressão Positiva/métodos , Insuficiência Respiratória/terapia , Volume de Ventilação Pulmonar/fisiologia , Idoso , Humanos , Inalação/fisiologia , Pessoa de Meia-Idade , Modelos Biológicos , Estudos Prospectivos , Insuficiência Respiratória/fisiopatologia , Fatores de Tempo , Resultado do TratamentoRESUMO
BACKGROUND AND OBJECTIVE: Healthcare datasets are plagued by issues of data scarcity and class imbalance. Clinically validated virtual patient (VP) models can provide accurate in-silico representations of real patients and thus a means for synthetic data generation in hospital critical care settings. This research presents a realistic, time-varying mechanically ventilated respiratory failure VP profile synthesised using a stochastic model. METHODS: A stochastic model was developed using respiratory elastance (Ers) data from two clinical cohorts and averaged over 30-minute time intervals. The stochastic model was used to generate future Ers data based on current Ers values with added normally distributed random noise. Self-validation of the VPs was performed via Monte Carlo simulation and retrospective Ers profile fitting. A stochastic VP cohort of temporal Ers evolution was synthesised and then compared to an independent retrospective patient cohort data in a virtual trial across several measured patient responses, where similarity of profiles validates the realism of stochastic model generated VP profiles. RESULTS: A total of 120,000 3-hour VPs for pressure control (PC) and volume control (VC) ventilation modes are generated using stochastic simulation. Optimisation of the stochastic simulation process yields an ideal noise percentage of 5-10% and simulation iteration of 200,000 iterations, allowing the simulation of a realistic and diverse set of Ers profiles. Results of self-validation show the retrospective Ers profiles were able to be recreated accurately with a mean squared error of only 0.099 [0.009-0.790]% for the PC cohort and 0.051 [0.030-0.126]% for the VC cohort. A virtual trial demonstrates the ability of the stochastic VP cohort to capture Ers trends within and beyond the retrospective patient cohort providing cohort-level validation. CONCLUSION: VPs capable of temporal evolution demonstrate feasibility for use in designing, developing, and optimising bedside MV guidance protocols through in-silico simulation and validation. Overall, the temporal VPs developed using stochastic simulation alleviate the need for lengthy, resource intensive, high cost clinical trials, while facilitating statistically robust virtual trials, ultimately leading to improved patient care and outcomes in mechanical ventilation.
Assuntos
Cuidados Críticos , Respiração Artificial , Humanos , Respiração Artificial/métodos , Estudos Retrospectivos , Simulação por Computador , Cuidados Críticos/métodos , Projetos de PesquisaRESUMO
BACKGROUND: Patient-specific respiratory mechanics models can support the evaluation of optimal lung protective ventilator settings during ventilation therapy. Clinical application requires that the individual's model parameter values must be identified with information available at the bedside. Multiple linear regression or gradient-based parameter identification methods are highly sensitive to noise and initial parameter estimates. Thus, they are difficult to apply at the bedside to support therapeutic decisions. METHODS: An iterative integral parameter identification method is applied to a second order respiratory mechanics model. The method is compared to the commonly used regression methods and error-mapping approaches using simulated and clinical data. The clinical potential of the method was evaluated on data from 13 Acute Respiratory Distress Syndrome (ARDS) patients. RESULTS: The iterative integral method converged to error minima 350 times faster than the Simplex Search Method using simulation data sets and 50 times faster using clinical data sets. Established regression methods reported erroneous results due to sensitivity to noise. In contrast, the iterative integral method was effective independent of initial parameter estimations, and converged successfully in each case tested. CONCLUSION: These investigations reveal that the iterative integral method is beneficial with respect to computing time, operator independence and robustness, and thus applicable at the bedside for this clinical application.
Assuntos
Modelos Biológicos , Mecânica Respiratória , Humanos , Análise de Regressão , Síndrome do Desconforto Respiratório/fisiopatologia , Fatores de TempoRESUMO
BACKGROUND: Mechanical ventilation (MV) is the primary form of support for acute respiratory distress syndrome (ARDS) patients. However, intra- and inter- patient-variability reduce the efficacy of general protocols. Model-based approaches to guide MV can be patient-specific. A physiological relevant minimal model and its patient-specific performance are tested to see if it meets this objective above. METHODS: Healthy anesthetized piglets weighing 24.0 kg [IQR: 21.0-29.6] underwent a step-wise PEEP increase manoeuvre from 5cmH2O to 20cmH2O. They were ventilated under volume control using Engström Care Station (Datex, General Electric, Finland), with pressure, flow and volume profiles recorded. ARDS was then induced using oleic acid. The data were analyzed with a Minimal Model that identifies patient-specific mean threshold opening and closing pressure (TOP and TCP), and standard deviation (SD) of these TOP and TCP distributions. The trial and use of data were approved by the Ethics Committee of the Medical Faculty of the University of Liege, Belgium. RESULTS AND DISCUSSIONS: 3 of the 9 healthy piglets developed ARDS, and these data sets were included in this study. Model fitting error during inflation and deflation, in healthy or ARDS state is less than 5.0% across all subjects, indicating that the model captures the fundamental lung mechanics during PEEP increase. Mean TOP was 42.4cmH2O [IQR: 38.2-44.6] at PEEP = 5cmH2O and decreased with PEEP to 25.0cmH2O [IQR: 21.5-27.1] at PEEP = 20cmH2O. In contrast, TCP sees a reverse trend, increasing from 10.2cmH2O [IQR: 9.0-10.4] to 19.5cmH2O [IQR: 19.0-19.7]. Mean TOP increased from average 21.2-37.4cmH2O to 30.4-55.2cmH2O between healthy and ARDS subjects, reflecting the higher pressure required to recruit collapsed alveoli. Mean TCP was effectively unchanged. CONCLUSION: The minimal model is capable of capturing physiologically relevant TOP, TCP and SD of both healthy and ARDS lungs. The model is able to track disease progression and the response to treatment.
Assuntos
Modelos Animais de Doenças , Pulmão/fisiologia , Pulmão/fisiopatologia , Síndrome do Desconforto Respiratório/fisiopatologia , Animais , Progressão da Doença , Modelos Biológicos , Ácido Oleico/efeitos adversos , Respiração com Pressão Positiva , Síndrome do Desconforto Respiratório/induzido quimicamente , Mecânica Respiratória/fisiologia , SuínosRESUMO
BACKGROUND: Spontaneous breathing (SB) effort during mechanical ventilation (MV) is an important metric of respiratory drive. However, SB effort varies due to a variety of factors, including evolving pathology and sedation levels. Therefore, assessment of SB efforts needs to be continuous and non-invasive. This is important to prevent both over- and under-assistance with MV. In this study, a machine learning model, Convolutional Autoencoder (CAE) is developed to quantify the magnitude of SB effort using only bedside MV airway pressure and flow waveform. METHOD: The CAE model was trained using 12,170,655 simulated SB flow and normal flow data (NB). The paired SB and NB flow data were simulated using a Gaussian Effort Model (GEM) with 5 basis functions. When the CAE model is given a SB flow input, it is capable of predicting a corresponding NB flow for the SB flow input. The magnitude of SB effort (SBEMag) is then quantified as the difference between the SB and NB flows. The CAE model was used to evaluate the SBEMag of 9 pressure control/ support datasets. Results were validated using a mean squared error (MSE) fitting between clinical and training SB flows. RESULTS: The CAE model was able to produce NB flows from the clinical SB flows with the median SBEMag of the 9 datasets being 25.39% [IQR: 21.87-25.57%]. The absolute error in SBEMag using MSE validation yields a median of 4.77% [IQR: 3.77-8.56%] amongst the cohort. This shows the ability of the GEM to capture the intrinsic details present in SB flow waveforms. Analysis also shows both intra-patient and inter-patient variability in SBEMag. CONCLUSION: A Convolutional Autoencoder model was developed with simulated SB and NB flow data and is capable of quantifying the magnitude of patient spontaneous breathing effort. This provides potential application for real-time monitoring of patient respiratory drive for better management of patient-ventilator interaction.