Rev Sci Instrum ; 83(10): 10D536, 2012 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-23130795

RESUMO

Enhancements to the JET poloidally scanning spectrometers are presented, which will aid the exploitation of the recently installed ITER-like wall in JET. They include the installation of visible filter∕photomultiplier tube assemblies and spectrometers and the replacement of large rotating mirrors in the JET vacuum with small oscillating mirrors outside. The upgrade has resulted in a more robust and reliable diagnostic than before, which is described. Drifts in the mirror angle reconstructed from quadrature encoder signals are found, a reference signal being required. The use of the small scanning mirrors necessitated the inclusion of focusing mirrors to maintain throughput into the vacuum ultraviolet spectrometers. The mirror design has taken account of the extreme sensitivity of the focusing to the grazing angle of incidence, an aspect of importance in the design of grazing incidence focusing components on future machines, such as ITER. The visible system has been absolutely calibrated using an in-vessel light source.

SELEÇÃO DE REFERÊNCIAS