RESUMO
AIMS: The GermAn Laser Lead Extraction RegistrY: GALLERY is a retrospective, national multicentre registry, investigating the safety and efficacy of laser lead extraction procedures in Germany. METHODS AND RESULTS: Twenty-four German centres that are performing laser lead extraction have participated in the registry. All patients, treated with a laser lead extraction procedure between January 2013 and March 2017, were consecutively enrolled. Safety and efficacy of laser lead extraction were investigated. A total number of 2524 consecutive patients with 6117 leads were included into the registry. 5499 leads with a median lead dwell time of 96 (62-141) months were treated. The mean number of treated leads per patient was 2.18 ± 1.02. The clinical procedural success rate was 97.86% and the complete lead removal was observed in 94.85%. Additional extraction tools were used in 6.65% of cases. The rate of procedural failure was 2.14% with lead age ≥10 years being its only predictor. The overall complication rate was 4.32%, including 2.06% major and 2.26% minor complications. Procedure-related mortality was 0.55%. Female sex and the presence of abandoned leads were predictors for procedure-related complications. The all-cause in-hospital mortality was 3.56% with systemic infection being the strongest predictor, followed by age ≥75 years and chronic kidney disease. CONCLUSION: In the GALLERY, a high success- and low procedure-related complication rates have been demonstrated. In multivariate analysis, female sex and the presence of abandoned leads were predictors for procedure-related complications, while the presence of systemic infection, age ≥75 years, and chronic kidney disease were independent predictors for all-cause mortality.
Assuntos
Desfibriladores Implantáveis , Marca-Passo Artificial , Insuficiência Renal Crônica , Idoso , Criança , Desfibriladores Implantáveis/efeitos adversos , Remoção de Dispositivo/métodos , Feminino , Humanos , Lasers de Excimer , Marca-Passo Artificial/efeitos adversos , Complicações Pós-Operatórias/etiologia , Sistema de Registros , Estudos Retrospectivos , Resultado do TratamentoRESUMO
In this article the case of a patient who received a total of 35 inappropriate defibrillator shocks due to insulation failure with externalized conductor of a Biotronik Linox® lead is described. The implanted defibrillator was immediately inactivated and the failed lead was extracted using a laser sheath system.
Assuntos
Desfibriladores Implantáveis/efeitos adversos , Traumatismos por Eletricidade/diagnóstico , Traumatismos por Eletricidade/etiologia , Falha de Equipamento , Traumatismos Cardíacos/diagnóstico , Traumatismos Cardíacos/etiologia , Idoso , Remoção de Dispositivo , Diagnóstico Diferencial , Traumatismos por Eletricidade/prevenção & controle , Eletrodos Implantados/efeitos adversos , Traumatismos Cardíacos/prevenção & controle , Humanos , Masculino , Traumatismo Múltiplo , Falha de TratamentoRESUMO
A case of pacemaker electrode infection with Staphylococcus aureus is presented. In order to avoid sternotomy, the right ventricular pacemaker lead, which had been implanted for 15 years, was successfully extracted using a laser sheath despite large endocarditic vegetations on the pacemaker lead. After completion of a 6-week course of antibiotics, the patient was discharged without any further evidence of infection.