1.
Lab Chip ; 9(9): 1282-5, 2009 May 07.
Artigo em Inglês | MEDLINE | ID: mdl-19370249

RESUMO

We have developed a novel, microfabricated, stand-alone microfluidic device that can efficiently sort and concentrate (bio-)analyte molecules by using kinesin motors and microtubules as a chemo-mechanical transduction machine. The device removes hundreds of targeted molecules per second from an analyte stream by translocating functionalized microtubules with kinesin across the stream and concentrating them at a horseshoe-shaped collector. Target biomolecule concentrations increase up to three orders of magnitude within one hour of operation.


Assuntos
Biopolímeros/química , Biopolímeros/isolamento & purificação , Técnicas Analíticas Microfluídicas/instrumentação , Técnicas Analíticas Microfluídicas/métodos , Proteínas Motores Moleculares/química , Proteínas Motores Moleculares/isolamento & purificação , Mapeamento de Interação de Proteínas/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Mapeamento de Interação de Proteínas/métodos