1.
Ophthalmology ; 121(2): 459-68, 2014 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-24289916

RESUMO

PURPOSE: To evaluate the performance of corneal hysteresis (CH), corneal resistance factor, and 16 investigator-derived Ocular Response Analyzer (ORA) variables in distinguishing keratoconus (KC) from the nondiseased state. DESIGN: Retrospective case series. PARTICIPANTS: Fifty-four eyes of 27 unaffected patients and 49 eyes of 25 KC patients from the Instituto de Olhos, Rio de Janeiro, Brazil. METHODS: Sixteen candidate variables were derived from exported ORA signals to characterize putative indicators of biomechanical behavior. Area under the receiver operating characteristic curve (AUC) and the Z statistic were used to compare diagnostic performance. MAIN OUTCOME MEASURES: Discriminant value of standard and derived ORA variables as measured by AUC. RESULTS: Fifteen of 16 candidate variables performed significantly better than chance (AUC, >0.5) at discriminating KC. Diagnostic performance was greatest for a custom variable related to the depth of deformation as defined by the minimum applanation signal intensity during corneal deformation (concavity(min); mean AUC ± standard error, 0.985 ± 0.002) and a new measure incorporating the pressure-deformation relationship of the entire response cycle (hysteresis loop area, 0.967 ± 0.002). Z statistics assessing the discriminative value of each of the top 5 variables demonstrated superiority to CH (AUC, 0.862 ± 0.002). Concavity(min) had the best overall predictive accuracy (cutoff value, 50.37; 94.9% sensitivity, 91.7% specificity, and 93.2% test accuracy), and the top 4 variables demonstrated the most consistent relationships to KC severity. CONCLUSIONS: Investigator-derived ORA variables related to the depth of deformation and the pressure-deformation relationship demonstrated very high test accuracy for detecting the presence of KC. Beyond their diagnostic value, the candidate variables described in this report provide mechanistic insight into the nature of the ORA signal and the characteristic changes in corneal dynamics associated with KC.


Assuntos
Córnea/fisiopatologia , Técnicas de Imagem por Elasticidade , Elasticidade/fisiologia , Ceratocone/diagnóstico , Adulto , Topografia da Córnea , Análise Discriminante , Feminino , Humanos , Pressão Intraocular , Ceratocone/fisiopatologia , Masculino , Curva ROC , Reprodutibilidade dos Testes , Estudos Retrospectivos , Sensibilidade e Especificidade
2.
Eye Contact Lens ; 40(6): 339-44, 2014 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25365551

RESUMO

PURPOSE: To assess changes in Ocular Response Analyzer (ORA) waveforms after UVA/riboflavin corneal collagen cross-linking (CXL) using investigator-derived and manufacturer-supplied morphometric variables in patients with keratoconus (KC) and postrefractive surgery ectasia. DESIGN: Prospective randomized trial of a standard epithelium-off CXL protocol. PARTICIPANTS: Patients with progressive KC (24 eyes of 21 patients) or postrefractive surgery ectasia (27 eyes of 23 patients) were enrolled. METHODS: Replicate ORA measurements were obtained before and 3 months after CXL. Pretreatment and posttreatment waveform variables were analyzed for differences by paired Student t tests using measurements with the highest waveform scores. MAIN OUTCOME MEASURES: Corneal hysteresis, corneal resistance factor, 37-s generation manufacturer-supplied ORA variables, and 15 investigator-derived ORA variables. RESULTS: No variables were significantly different 3 months after CXL in the KC group, and no manufacturer-supplied variables changed significantly in the postrefractive surgery ectasia group. Four custom variables (ApplanationOnsetTime, P1P2avg, Impulse, and Pmax) increased by small but statistically significant margins after CXL in the postrefractive surgery ectasia group. CONCLUSIONS: Changes in a small subset of investigator-derived variables suggested an increase in corneal bending resistance after CXL. However, the magnitudes of these changes were low and not commensurate with the degree of clinical improvement or prior computational estimates of corneal stiffening in the same cohort over the same period. Available air-puff-derived measures of the corneal deformation response underestimate the biomechanical changes produced by CXL.


Assuntos
Fenômenos Biomecânicos/fisiologia , Doenças da Córnea/tratamento farmacológico , Reagentes de Ligações Cruzadas/uso terapêutico , Ceratocone/tratamento farmacológico , Fotoquimioterapia , Adulto , Colágeno/efeitos dos fármacos , Córnea/efeitos dos fármacos , Doenças da Córnea/etiologia , Doenças da Córnea/fisiopatologia , Topografia da Córnea , Dilatação Patológica/tratamento farmacológico , Dilatação Patológica/fisiopatologia , Feminino , Humanos , Pressão Intraocular/fisiologia , Ceratocone/fisiopatologia , Masculino , Pessoa de Meia-Idade , Fármacos Fotossensibilizantes/uso terapêutico , Estudos Prospectivos , Riboflavina/uso terapêutico , Raios Ultravioleta , Acuidade Visual/fisiologia
3.
Am J Ophthalmol ; 173: 16-22, 2017 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-27702622

RESUMO

PURPOSE: To correlate intraoperative interface fluid dynamics during Descemet stripping automated endothelial keratoplasty (DSAEK) using intraoperative optical coherence tomography (iOCT) in the Prospective Intraoperative and Perioperative Ophthalmic Imaging with Optical Coherence Tomography (PIONEER) study with postoperative outcomes. DESIGN: Prospective consecutive, interventional, comparative case series. PARTICIPANTS: One hundred seventy-eight eyes of 173 patients undergoing DSAEK from the Cole Eye Institute, Cleveland, Ohio. METHODS: Eyes that underwent DSAEK between October 2011 and March 2014 from the PIONEER intraoperative and perioperative OCT study were included. An automated interface fluid segmentation algorithm evaluated intraoperative dynamics of interface fluid before and after surgical manipulations. iOCT images were also captured at multiple intraoperative time points for 2 different DSAEK techniques, 1 that used an active air infusion system and 1 that did not. MAIN OUTCOME MEASURES: Interface fluid metrics, graft nonadherence. RESULTS: iOCT measurements of interface fluid after final surgical manipulations and immediately before leaving the operating room identified that total fluid volume (P = .002), largest fluid volume pocket (P = .002), max fluid area (P = .006), mean fluid thickness (P = .03), and max fluid thickness (P = .01) significantly correlated with graft nonadherence rates within the first postoperative week. After placement and optimization of intraoperative lenticle adherence, iOCT revealed a significant difference between the area, volume, and thickness of maximum fluid pockets between the 2 surgical techniques, but both techniques resulted in significant reduction of interface fluid during the procedure. CONCLUSIONS: Larger residual interface fluid volume, area, and thickness at the end of surgery detected with iOCT are associated with early graft nonadherence and can be quantified with an automated algorithm. iOCT imaging can successfully capture technique-dependent differences in fluid dynamics during DSAEK.


Assuntos
Ceratoplastia Endotelial com Remoção da Lâmina Limitante Posterior/métodos , Endotélio Corneano/transplante , Distrofia Endotelial de Fuchs/cirurgia , Monitorização Intraoperatória/métodos , Cirurgia Assistida por Computador/métodos , Tomografia de Coerência Óptica/métodos , Idoso , Feminino , Seguimentos , Distrofia Endotelial de Fuchs/diagnóstico , Humanos , Hidrodinâmica , Masculino , Estudos Prospectivos , Resultado do Tratamento
4.
J Refract Surg ; 32(7): 479-94, 2016 Jul 01.
Artigo em Inglês | MEDLINE | ID: mdl-27400080

RESUMO

PURPOSE: To evaluate the performance of the Ocular Response Analyzer (ORA) (Reichert Ophthalmic Instruments, Depew, NY) variables and Pentacam HR (Oculus Optikgeräte GmbH, Wetzlar, Germany) tomographic parameters in differentiating forme fruste keratoconus (FFKC) from normal corneas, and to assess a combined biomechanical and tomographic parameter to improve outcomes. METHODS: Seventy-six eyes of 76 normal patients and 21 eyes of 21 patients with FFKC were included in the study. Fifteen variables were derived from exported ORA signals to characterize putative indicators of biomechanical behavior and 37 ORA waveform parameters were tested. Sixteen tomographic parameters from Pentacam HR were tested. Logistic regression was used to produce a combined biomechanical and tomography linear model. Differences between groups were assessed by the Mann-Whitney U test. The area under the receiver operating characteristics curve (AUROC) was used to compare diagnostic performance. RESULTS: No statistically significant differences were found in age, thinnest point, central corneal thickness, and maximum keratometry between groups. Twenty-one parameters showed significant differences between the FFKC and control groups. Among the ORA waveform measurements, the best parameters were those related to the area under the first peak, p1area1 (AUROC, 0.717 ± 0.065). Among the investigator ORA variables, a measure incorporating the pressure-deformation relationship of the entire response cycle was the best predictor (hysteresis loop area, AUROC, 0.688 ± 0.068). Among tomographic parameters, Belin/Ambrósio display showed the highest predictive value (AUROC, 0.91 ± 0.057). A combination of parameters showed the best result (AUROC, 0.953 ± 0.024) outperforming individual parameters. CONCLUSIONS: Tomographic and biomechanical parameters demonstrated the ability to differentiate FFKC from normal eyes. A combination of both types of information further improved predictive value. [J Refract Surg. 2016;32(7):479-485.].


Assuntos
Córnea/fisiopatologia , Técnicas de Diagnóstico Oftalmológico , Elasticidade/fisiologia , Ceratocone/diagnóstico , Adolescente , Adulto , Área Sob a Curva , Fenômenos Biomecânicos , Criança , Paquimetria Corneana , Topografia da Córnea , Feminino , Humanos , Ceratocone/fisiopatologia , Masculino , Pessoa de Meia-Idade , Curva ROC , Estudos Retrospectivos , Tomografia , Adulto Jovem
5.
Am J Ophthalmol ; 164: 14-21, 2016 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-26743618

RESUMO

PURPOSE: To evaluate the performance of corneal hysteresis (CH), corneal resistance factor (CRF), 37 Ocular Response Analyzer (ORA) waveform parameters, and 15 investigator-derived ORA variables in differentiating forme fruste keratoconus (KC) from normal corneas. DESIGN: Case-control study. METHODS: Seventy-eight eyes of 78 unaffected patients and 21 topographically normal eyes of 21 forme fruste KC patients with topographically manifest KC in the contralateral eye were matched for age, the thinnest point of the cornea, central corneal thickness, and maximum keratometry. Fifteen candidate variables were derived from exported ORA signals to characterize putative indicators of biomechanical behavior, and 37 waveform parameters were tested. Differences between groups were assessed by the Mann-Whitney test. The area under the receiver operating characteristic curve (AUROC) was used to compare the diagnostic performance. RESULTS: Ten of 54 parameters reached significant differences between the groups (Mann-Whitney test, P < .05). Neither CRF nor CH differed significantly between the groups. Among the ORA waveform measurements, the best parameters were those related to the area under the first peak, p1area, and p1area1 (AUROC, 0.714 ± 0.064 and 0.721 ± 0.065, respectively). Among the investigator ORA variables, a measure incorporating the pressure-deformation relationship of the entire response cycle performed best (hysteresis loop area, AUROC, 0.694 ± 0.067). CONCLUSION: Waveform-derived ORA parameters, including a custom measure incorporating the pressure-deformation relationship of the entire response cycle, performed better than traditional CH and CRF parameters in differentiating forme fruste KC from normal corneas.


Assuntos
Córnea/fisiopatologia , Elasticidade/fisiologia , Ceratocone/diagnóstico , Adulto , Área Sob a Curva , Astigmatismo/fisiopatologia , Fenômenos Biomecânicos , Estudos de Casos e Controles , Paquimetria Corneana , Topografia da Córnea , Técnicas de Diagnóstico Oftalmológico , Dilatação Patológica/diagnóstico , Dilatação Patológica/fisiopatologia , Análise Discriminante , Feminino , Humanos , Ceratocone/fisiopatologia , Masculino , Curva ROC , Adulto Jovem
6.
J Cataract Refract Surg ; 40(6): 918-28, 2014 06.
Artigo em Inglês | MEDLINE | ID: mdl-24726160

RESUMO

PURPOSE: To analyze changes in new biomechanical descriptors with myopic femtosecond laser-assisted laser in situ keratomileusis (LASIK), compare them with the biomechanical response after photorefractive keratectomy (PRK) with similar levels of myopic ablation, and evaluate correlations between changes in custom variables and biomechanically relevant variables. SETTING: Cleveland Clinic, Cleveland, Ohio, USA. DESIGN: Cohort study. METHODS: Custom biomechanical variables from the Optical Response Analyzer were assessed preoperatively and 1 and 3 months postoperatively. Differences between preoperative values and postoperative values were determined. Intraindividual change (preoperative value minus postoperative value) was calculated and compared with changes after PRK. The correlation of the change in each custom biomechanical variable with the preoperative central corneal thickness, residual stromal bed tissue ablated, and percentage of tissue depth altered was also studied. RESULTS: The study enrolled 156 eyes of 156 consecutive patients. Fifteen variables changed significantly after femtosecond myopic LASIK and were stable postoperatively because no significant difference was shown between 1-month values and 3-month values. Comparison of the changes in biomechanical variables between LASIK and PRK eyes showed no significant differences. Surgical changes in several custom biomechanical variables correlated with the percentage of tissue depth altered. CONCLUSIONS: The results provide the first reference values for a more comprehensive panel of indicators of the biomechanical response to myopic LASIK and PRK. Changes in custom variables reflected a consistent decrease in corneal biomechanical resistance to deformation after myopic femtosecond LASIK and PRK. For comparable attempted corrections, biomechanical changes were comparable between femtosecond laser-assisted LASIK and PRK. FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosures are listed after the references.


Assuntos
Córnea/fisiopatologia , Elasticidade/fisiologia , Ceratomileuse Assistida por Excimer Laser In Situ/métodos , Lasers de Excimer/uso terapêutico , Miopia/cirurgia , Ceratectomia Fotorrefrativa/métodos , Adulto , Fenômenos Biomecânicos , Estudos de Coortes , Paquimetria Corneana , Feminino , Humanos , Masculino , Miopia/fisiopatologia , Estudos Prospectivos